Hui Guo, Jiantao Li, Yuan Gao, Shaobo Nie, Chenliang Quan, Jia Li, Wei Zhang, "A Finite Element Study on the Treatment of Thoracolumbar Fracture with a New Spinal Fixation System", BioMed Research International, vol. 2021, Article ID 8872514, 9 pages, 2021. https://doi.org/10.1155/2021/8872514

A Finite Element Study on the Treatment of Thoracolumbar Fracture with a New Spinal Fixation System

Academic Editor: Frederic CAILOTTO
Received13 Sep 2020
Revised29 Nov 2020
Accepted25 Mar 2021
Published12 Apr 2021

Abstract

Objective. In this study, the mechanical properties of the new spinal fixation system (NSFS) in the treatment of thoracolumbar fractures were evaluated by the finite element analysis method, so as to provide a mechanical theoretical basis for the later biomechanical experiments and clinical experiments. Methods. T12-L2 bone model was constructed to simulate L1 vertebral fracture, and three models of internal fixation systems were established on the basis of universal spinal system (USS): Model A: posterior short-segment fixation including the fractured vertebra (PSFFV); Model B: short-segment pedicle screw fixation (SSPF); Model C: new spinal fixation system (NSFS). After assembling the internal fixation system and fracture model, the finite element analysis was carried out in the ANSYS Workbench 18.0 software, and the stress of nail rod system, fracture vertebral body stress, vertebral body mobility, and vertebral body displacement were recorded in the three models. Results. The peak values of internal fixation stress, vertebral body stress, vertebral body maximum displacement, and vertebral body maximum activity in Model C were slightly smaller than those in Model B. Conclusions. Compared with the traditional internal fixation system, the new spinal internal fixation system may have the mechanical advantage and can provide sufficient mechanical stability for thoracolumbar fractures.

1. Introduction

Spinal fractures account for 5%-6% of systemic fractures, and thoracolumbar fractures are the most common [1, 2]. The lumbar spine is more susceptible to the load-bearing force than the thoracic and cervical spine [3]. Stability is an important factor in the treatment of vertebral column fractures. The occurrence of burst fractures is due to axial compression and damage to the anterior and middle columns, which may lead to instability and spinal cord injury. In the past thirty years, surgical management has made tremendous progress, including various surgical stabilization strategy [4, 5]. Pedicle screw-rod system internal fixation is the most commonly used surgical method for the treatment of thoracolumbar fractures, including fracture reduction, correction of kyphosis, relief of nerve compression, and reconstruction of spinal stability. It can achieve fracture reduction, correct kyphosis, relieve nerve compression, and reconstruct spine stability [68]. The reduction mechanism of the pedicle screw-rod system is to restore the height of the fractured vertebral body through indirect traction, so as to correct kyphosis. The indirect reduction method cannot restore the height of fracture vertebral body and kyphotic deformity to a limited extent and cannot fully anatomize the reduction. After reduction, a cavity will be formed in the vertebral body, resulting in insufficient front support and further loss of vertebral body height in the long term. In some patients, kyphosis even worsened, and residual back pain or even failure of internal fixation occurred [9, 10].

In order to solve the problems of poor reduction of traditional instruments and insufficient anterior support, the author designed a new set of spinal internal fixation instruments, which include intravertebral distraction reduction device, minimally invasive pedicle bone graft supporting nail, vertebral implant, connector, and fixation system. In theory, this set of instruments can achieve anatomical reduction, adequate anterior support, and sufficient stability of the spine and, finally, solve the problem of loss of vertebral height in long-term fractures and avoid permanent low back pain or internal fixation failure. In order to prove the mechanical advantages of the new internal fixation system, this study compares the mechanical differences between the NSFS and the traditional internal fixation system in the treatment of thoracolumbar fractures by means of finite element analysis. The purpose of this study involved exploring the advantages and disadvantages of NSFS, PSFFV, and SSPF for the fixation of L1 vertebral fracture by using finite element analysis. The report is as follows.

2. Methods

2.1. Establishment of Finite Element Model

The CT images of thoracolumbar vertebrae were imported into the three-dimensional reconstruction software mimics 20 (Materialise Company, Belgium) in DICOM format, and the three vertebral models of T12 L1 and L2 were reconstructed. Import the STL file format into Geomagic 2012 (Geomagic, USA) to build a curved surface model and perform V-shaped osteotomy on L1 vertebrae to construct a fracture model (Figure 1).

Using SolidWorks 2015 (Dassault, France) establishes a three-dimensional model of pedicle screw rod system and assembles the pedicle screw rod system and vertebral model. Three fixation models and postassembly models of pedicle screw rod system is constructed, as shown in Figure 2.

The bone and the intervertebral disc meshed into eight-node hexahedral elements with Solid 187 with a mesh size of 1 mm were chosen. There are 436283 and 711037 units in Model A, 459409 nodes and 750985 units in Model B, and 493072 nodes and 809052 cell units in Model C. The assembled entity model is imported into the software ANSYS Workbench 18.0 (ANSYS company, USA) for Boolean operation analysis. Units and nodes of model were shown in Table 1.


ModelUnitesNodes

A436283711037
B459409750985
C493072809052

2.2. Material Properties, Boundary Conditions, and Loading

According to the attachment positions, the spinal ligaments were established (Figure 3) and defined as nonlinear hyperplastic material using Combine 39, which only allowed for tension deformation without compression behavior. The mechanical behaviors of the spinal ligaments were based on nonlinear stress-strain curves [11]. The nonlinear properties of ligaments were shown in Table 2. The elastic properties and Poisson’s ratio of various structural materials were shown in Table 3 [11, 12].


ALLPLLFLISLSSLITLCL
D (mm)F (N)D (mm)F (N)D (mm)F (N)D (mm)F (N)D (mm)F (N)D (mm)F (N)D (mm)F (N)

-500-500-500-500-500-500-500
00000000000000
0.2317311.5010.935836.05333.64194.87811.29524.26831.9814.87813.7796.09761.89846.0533
0.4010722.3971.02512.1074.388623.1711.676214.0242.483820.7324.967610.9762.415313.317
0.561543.5841.158732.0824.784840.8541.874323.1712.8839.6345.561922.5613.05739.952
0.6773665.3751.256755.0855.074362.8052.026741.4633.154359.7566.095242.6833.279961.138
0.81105106.541.4617148.315.379107.322.163861.5853.6419103.056.552464.6343.6096105.93
1.016200.971.5865248.795.6076151.222.3771104.884.0076151.227.070599.393.877151.33
1.0695249.45.8971249.392.5448158.544.2971204.887.5886146.954.0374202.79
2.6514209.154.5105249.397.92382004.1533249.4
2.6971249.398.1676248.78

ALL: anterior longitudinal ligament; PLL: posterior longitudinal ligament; FL: flaval ligament; ISL: interspinous ligament; SSL: supraspinous ligament; ITL: intertransverse ligament; CL: capsular ligaments; D: deflection; F: force.

MaterialYoung modulus E (MPa)V Poisson ratio

Cortical bone/endplate120000.3
Cancellous bone1000.2
Posterior arch35000.25
Annulus fibrosus4.20.4
Nucleus pulposus10.49
Spinal instrumentation1100000.3

The contact relationship was set as follows. The interfaces between the screw and bone were set to be bonded. The cortical bone and cancellous bone were set to be bonded. The vertebral body was bound to the adjacent intervertebral disc. And nucleus pulposus and annulus fibrosus were set to be bonded.

In order to simulate the real stress of human lumbar vertebrae, the lower endplate of L2 was restrained and fixed. 500 N axial compression force was applied to the upper endplate of T12 of three groups of models, and pure bending moment of 7.5 Nm was applied to all directions at the same time.

2.3. Evaluation Index

The evaluation indexes include values and distribution of internal fixation stress of three finite element models under flexion, extension, right flexion, and right rotation loading conditions; fracture vertebral body stress; vertebral range of motion (ROM); and vertebral displacement.

3. Results

3.1. Stress Comparison of Internal Fixation System

The peak value of stress shows that Model A has the largest peak value of stress, which is 210 MPa for forward bending, 145 MPa for backward extension, 262 MPa for right flexion, and 282 MPa for right rotation. The comparison results of the peak values of stress in three Models are A > B > C for forward bending, backward extension, and right rotation and A > C > B for right flexion (Figure 4).

The stress distribution nephogram shows that the stress distribution of the nail bar in Model A is more concentrated, mainly on the root of the screw and the rod above the cross-link; there is no obvious stress concentration in Model B and C. Stress distribution is more scattered, the two screws added in the middle share part of the stress, and the stress on the rod is evenly distributed on the whole (Figure 5).

3.2. Comparison of Fracture Vertebral Body Stress

Comparing the peak stress of the fractured vertebral body, the peak stress of Model A is the largest, which is 167 MPa of flexion, 62 MPa of posterior extension, 126 MPa of right flexion, and 113 MPa of right rotation. The comparison results of the peak stress of the three groups are flexion, extension, right flexion, and right rotation under all four working conditions: A > B > C (Figure 6).

The stress distribution nephogram of vertebral body shows that the stress is mainly concentrated on the upper endplate and the middle column, especially the apex of “V”-shaped osteotomy. Compared with the stress distribution transmitted to the lower end plate, the stress distribution of the lower end plate in Model A was the most scattered, which may be related to the screw support in the fracture plane of Model B and C (Figure 7).

3.3. Comparison of Vertebral Displacement and Mobility

Vertebral displacement showed that Model A had the largest vertebral displacement, which is 2.1 mm for forward flexion, 0.8 mm for posterior extension, 1.9 mm for right flexion, and 1.8 mm for right circumflexion. The comparison results of vertebral displacement of the three Models were flexion forward and right flexion A > B = C; the posterior extension and right rotation A > B > C.

The ROM of vertebral body in Model A was 3.10° in flexion, 1.76° in extension, 2.90° in right flexion, and 2.81° in right rotation. The comparison of ROM among the three Models was as follows: flexion, extension, right flexion, and right rotation A > B > C. (Figure 8).

4. Discussion

Thoracolumbar fractures are often treated with posterior short segment fixation [1315]. There are two traditional short segment fixation methods: posterior short-segment fixation including the fractured vertebra (PSFFV) and short-segment pedicle screw fixation (SSPF) [16]. In this study, by means of finite element simulation mechanical analysis, we compared the mechanical differences among PSFFV (Model A), traditional SSPF (Model B), and NSFS (Model C). The results showed that the NSFS had significant mechanical advantages in the fixation of vertebral fractures.

Comparison between Model C and Model A, Internal fixation stress: the stress distribution of the internal fixation system in Model C is more scattered, and the peak value of stress is smaller than that in Model A, which indicates that the fracture risk of the internal fixation system in Model C is lower than that in Model A. Spine stability: the vertebral body stress, vertebral displacement, and ROM of the internal fixation system in Model C were lower than those in Model A, which proved that the spinal stability of the internal fixation system in Model C was stronger. Through the study and comparison, we think that the NSFS has obvious mechanical advantages over the PSFFV, mainly because the new internal fixation system increases the plane nail placement of the injury vertebra, which makes the spine more stable, the stress more scattered, and reduces the risk of breaking the nail and the rod. This is consistent with the research results of Liao et al. [17], who found that for vertebral fracture, the fixation method of SSPF has more stable fixation strength and more scattered stress than that of PSFFV. In a biomechanical study of fresh frozen cadaver, Baaj et al. [18] found that compared with fixation across injured vertebrae, fixation of injured vertebrae can increase structural stability by 25%, and also has the quadrilateral effect and suspension effect to avoid the fixation of cross segments. It has the mechanical advantages of dispersing and reducing the internal fixation stress and significantly reducing the failure of internal fixation. Elbehairy et al. [19] found that 32 patients with thoracolumbar fractures fixed by SSPF can provide more stable vertebral fixation strength.

Comparison between Model C and Model B, Internal fixation stress: the distribution of internal fixation stress in the two models is relatively similar, which is relatively scattered. Except that the stress peak value of Model C was greater than that of Model B under the right flexion condition. The stress peak value of Model C was lower than that of Model B under other conditions, indicating that the fracture risk of the NSFS of Model C was lower. Spine stability: the vertebral body displacement and vertebral body ROM of the internal fixation system in Model Care were less than that in Model B, which proves that the spinal stability in Model C is stronger than that in Model B. Through research and comparison, we believe that the NSFS has significant mechanical advantages over the traditional SSPF, which may be related to the way the screw is placed in the injured vertebral plane. The two screws enter into the vertebral body through the middle point of the outer wall of the pedicle root, respectively, so as to provide more cortical bone fixation, increase the cohesion angle of the screw placement, make the screw placement longer, and increase the stability of the spine. Dvorak et al. [20] compared pedicle internal fixation with pedicle external fixation, proving that pedicle external fixation can provide better spinal stability; Klane et al. [21] proved that there was little difference in biomechanics between pedicle internal screw and pedicle external screw through biomechanical experiments, through the study of 8 fresh cadavers. Barber et al. [22] found that the greater the cohesion angle of pedicle screws, the more resistant it was to pull out stress and provided better spinal stability. In addition, we found that the NSFS is not collinear (the middle injuried vertebra is outside), which provides multiplane fixation control, thus increasing its stability against loads in different directions.

Besides, the research shows that there are some shortcomings in the traditional fixation system: (1) insufficient fracture reduction and (2) insufficient anterior support [2325]. (3) All lumbar paraspinal muscles were rendered useless, which may be another major limitation of this study. The traditional SSPF system replaces the fractured vertebral body by indirect traction. The fractured vertebral body can be effectively reduced by the traction of the fibrous annulus of the intervertebral disc, but the central bone mass of the anterior and middle column cannot be pulled, resulting in a central collapse defect, resulting in insufficient reduction [26, 27]. Although the height and shape of the fractured vertebral body are restored after indirect reduction, it is difficult to restore the trabecular structure in the vertebral body. The traditional fixation method of injured vertebrae is to place nails before reduction, which occupies the channels of reduction and bone grafting in the vertebral body, thus cannot be further operated in the injured vertebrae, resulting in a cavity in the vertebral body and insufficient anterior support [28]. The NSFS designed in this study can solve the above problems, and its principle is as follows: first, it is the same as the internal fixation system of cross injured vertebrae, and it is indirectly braced and fixed with cross injured vertebrae nails; then, through the pedicle passage at the midpoint of the transverse process root of injured vertebra, a new type of intravertebral distraction device was used for direct distraction reduction in the vertebral body and anatomical reduction of the fractured vertebral body. Then, the vertebral cavity was filled with bone graft to provide anterior support of vertebral body, and finally, the nails were placed on both sides. After connecting the rods, NSFS were combined into a whole. It realizes the anatomical reduction of the fractured vertebral body, provides sufficient anterior support of injured vertebra, and ensures the stability of the spine at the same time.

This study is also limited by this study is only a finite element analysis. Stress and displacement of the spine were investigated as there are considered important varies. The biomechanical propriety was well described in this study but not all biomechanical parameters. The further biomechanical experiments need to be performed to prove the results of this study.

5. Conclusion

The results of finite element mechanical analysis show that the new spinal internal fixation system may have mechanical advantages over the traditional internal fixation system in the treatment of vertebral fracture, which can provide sufficient mechanical stability at the fracture end, and it is worth further promotion and application.

Abbreviations

PSFFV:Posterior short-segment fixation including the fractured vertebra
SSPF:Short-segment pedicle screw fixation
NSFS:New spinal fixation system
USS:Universal spinal system
USA:United States of America
ROM:Range of motion.

Data Availability

Data are available from the corresponding author upon request.

Conflicts of Interest

Each author certifies that he has no commercial associations that might pose a conflict of interest with the submitted article.

Authors’ Contributions

Wei Zhang did the conception and design. Hui Guo and Jiantao Li performed the acquisition of data. Hui Guo and Yuan Gao performed the analysis and interpretation of data. Wei Zhang critically revised the article. All authors reviewed the submitted version of the manuscript. Wei Zhang approved the final version of the manuscript on behalf of all authors. Wei Zhang and Jia Li supervised the study. Hui Guo, Jiantao Li, and Yuan Gao contributed equally to this work.

Acknowledgments

This study was supported by the Transformation project of Chinese PLA General Hospital (ZH19006).

References

  1. P. Heidari, M. R. Zarei, M. R. Rasouli, A. R. Vaccaro, and V. Rahimi-Movaghar, “Spinal fractures resulting from traumatic injuries,” Chinese journal of traumatology = Zhonghua chuang shang za zhi, vol. 13, no. 1, pp. 3–9, 2010. View at: Google Scholar
  2. Z. Dengfeng, W. Haojie, and W. Xiao, “Comparative study of the clinical effect and safety of anterior surgical approach and posterior surgical approach in the treatment of thoracolumbar spinal fracture,” Open Medicine, vol. 10, no. 1, pp. 410–415, 2015. View at: Publisher Site | Google Scholar
  3. B. Ozdemir, A. Kanat, O. Batcik, H. Gucer, and C. Yolas, “Ligamentum flavum hematomas: why does it mostly occur in old Asian males? Interesting point of reported cases: review and case report,” Journal of Craniovertebral Junction & Spine, vol. 7, no. 1, pp. 7–12, 2016. View at: Publisher Site | Google Scholar
  4. A. Kanat and B. Ozdemir, “In reply to the letter to the editor regarding “restoration of anterior vertebral height by short-segment pedicle screw fixation with screwing of fractured vertebra for the treatment of unstable thoracolumbar fractures”,” World Neurosurgery, vol. 101, p. 793, 2017. View at: Publisher Site | Google Scholar
  5. B. Ozdemir, A. Kanat, C. Erturk et al., “Restoration of anterior vertebral height by short-segment pedicle screw fixation with screwing of fractured vertebra for the treatment of unstable thoracolumbar fractures,” World Neurosurgery, vol. 99, pp. 409–417, 2017. View at: Publisher Site | Google Scholar
  6. J. K. Scheer, J. Bakhsheshian, S. Fakurnejad, T. Oh, N. S. Dahdaleh, and Z. A. Smith, “Evidence-based medicine of traumatic thoracolumbar burst fractures: a systematic review of operative management across 20 years,” Global Spine Journal, vol. 5, no. 1, pp. 73–82, 2015. View at: Publisher Site | Google Scholar
  7. T. Tan, M. S. Huang, J. Mathew et al., “Anterior versus posterior approach in the management of AO Type B1 & B2 traumatic thoracolumbar fractures: a level 1 trauma centre study,” Journal of Clinical Neuroscience, vol. 72, pp. 219–223, 2020. View at: Publisher Site | Google Scholar
  8. M. Cahueque, A. Cobar, C. Zuñiga, and G. Caldera, “Management of burst fractures in the thoracolumbar spine,” Journal of Orthopaedics, vol. 13, no. 4, pp. 278–281, 2016. View at: Publisher Site | Google Scholar
  9. C. Mazel and L. Ajavon, “Malunion of post-traumatic thoracolumbar fractures,” Orthopaedics & Traumatology, Surgery & Research: OTSR., vol. 104, no. 1, pp. S55–S62, 2018. View at: Publisher Site | Google Scholar
  10. H. L. Ren, J. X. Wang, and J. M. Jiang, “Is short same-segment fixation really better than short-segment posterior fixation in the treatment of thoracolumbar fractures?” Spine, vol. 43, no. 21, pp. 1470–1478, 2018. View at: Publisher Site | Google Scholar
  11. H. Schmidt, F. Heuer, J. Drumm, Z. Klezl, L. Claes, and H. Wilke, “Application of a calibration method provides more realistic results for a finite element model of a lumbar spinal segment,” Clinical Biomechanics, vol. 22, no. 4, pp. 377–384, 2007. View at: Publisher Site | Google Scholar
  12. C. Li, Y. Zhou, H. Wang, J. Liu, and L. Xiang, “Treatment of unstable thoracolumbar fractures through short segment pedicle screw fixation techniques using pedicle fixation at the level of the fracture: a finite element analysis,” PloS One, vol. 9, no. 6, article e99156, 2014. View at: Publisher Site | Google Scholar
  13. J. J. Chokshi and M. Shah, “Outcomes of including fracture level in short-segment fixation for Th oracolumbar fracture dislocation,” Asian Spine Journal, vol. 13, no. 1, pp. 56–60, 2019. View at: Publisher Site | Google Scholar
  14. T. A. Aly, “Short segment versus long segment pedicle screws fixation in management of thoracolumbar burst fractures: meta-analysis,” Asian Spine Journal, vol. 11, no. 1, pp. 150–160, 2017. View at: Publisher Site | Google Scholar
  15. A. F. Joaquim, J. P. Maslak, and A. A. Patel, “Spinal reconstruction techniques for traumatic spinal injuries: a systematic review of biomechanical studies,” Global Spine Journal, vol. 9, no. 3, pp. 338–347, 2019. View at: Publisher Site | Google Scholar
  16. M. Cimatti, S. Forcato, F. Polli, M. Miscusi, A. Frati, and A. Raco, “Pure percutaneous pedicle screw fixation without arthrodesis of 32 thoraco-lumbar fractures: clinical and radiological outcome with 36-month follow-up,” European Spine Journal, vol. 22, Suppl 6, pp. 925–932, 2013. View at: Publisher Site | Google Scholar
  17. J. C. Liao, W. P. Chen, and H. Wang, “Treatment of thoracolumbar burst fractures by short-segment pedicle screw fixation using a combination of two additional pedicle screws and vertebroplasty at the level of the fracture: a finite element analysis,” BMC Musculoskeletal Disorders, vol. 18, no. 1, p. 262, 2017. View at: Publisher Site | Google Scholar
  18. A. A. Baaj, P. M. Reyes, A. S. Yaqoobi et al., “Biomechanical advantage of the index-level pedicle screw in unstable thoracolumbar junction fractures,” Journal of Neurosurgery. Spine, vol. 14, no. 2, pp. 192–197, 2011. View at: Publisher Site | Google Scholar
  19. H. F. el Behairy, A. M. Abdelaziz, A. K. Saleh et al., “Short-segment fixation of thoracolumbar fractures with incorporated screws at the level of fracture,” Orthopaedic Surgery, vol. 12, no. 1, pp. 170–176, 2020. View at: Publisher Site | Google Scholar
  20. M. Dvorak, S. MacDonald, K. R. Gurr, S. I. Bailey, and R. G. Haddad, “An anatomic, radiographic, and biomechanical assessment of extrapedicular screw fixation in the thoracic spine,” Spine, vol. 18, no. 12, pp. 1689–1694, 1993. View at: Publisher Site | Google Scholar
  21. K. K. White, R. Oka, A. T. Mahar, A. Lowry, and S. R. Garfin, “Pullout strength of thoracic pedicle screw instrumentation: comparison of the transpedicular and extrapedicular techniques,” Spine, vol. 31, no. 12, pp. E355–E358, 2006. View at: Publisher Site | Google Scholar
  22. J. W. Barber, S. D. Boden, T. Ganey, and W. C. Hutton, “Biomechanical study of lumbar pedicle screws: does convergence affect axial pullout strength?” Journal of Spinal Disorders, vol. 11, no. 3, pp. 215–220, 1998. View at: Google Scholar
  23. R. M. Kanna, A. P. Shetty, and S. Rajasekaran, “Posterior fixation including the fractured vertebra for severe unstable thoracolumbar fractures,” The Spine Journal, vol. 15, no. 2, pp. 256–264, 2015. View at: Publisher Site | Google Scholar
  24. H. Aono, K. Ishii, H. Tobimatsu et al., “Temporary short-segment pedicle screw fixation for thoracolumbar burst fractures: comparative study with or without vertebroplasty,” The Spine Journal, vol. 17, no. 8, pp. 1113–1119, 2017. View at: Publisher Site | Google Scholar
  25. F. De Iure, G. Lofrese, P. De Bonis, F. Cultrera, M. Cappuccio, and S. Battisti, “Vertebral body spread in thoracolumbar burst fractures can predict posterior construct failure,” The Spine Journal, vol. 18, no. 6, pp. 1005–1013, 2018. View at: Publisher Site | Google Scholar
  26. U. J. Spiegl, A. B. Ahrberg, C. Anemüller et al., “Which anatomic structures are responsible for the reduction loss after hybrid stabilization of osteoporotic fractures of the thoracolumbar spine?” BMC Musculoskeletal Disorders, vol. 21, no. 1, p. 54, 2020. View at: Publisher Site | Google Scholar
  27. M. L. Pachowsky, A. Kleyer, L. Wegener et al., “Quantitative T2 mapping shows increased degeneration in adjacent intervertebral discs following kyphoplasty,” Cartilage, vol. 11, no. 2, pp. 152–159, 2020. View at: Publisher Site | Google Scholar
  28. H. Hu, X. Lin, L. Tan et al., “Imaging study of “shell” phenomenon of thoracolumbar fractures after posterior reduction,” Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese Journal of Reparative and Reconstructive Surgery, vol. 31, no. 8, pp. 976–981, 2017. View at: Publisher Site | Google Scholar

Copyright © 2021 Hui Guo et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Related articles

No related content is available yet for this article.
 PDF Download Citation Citation
 Download other formatsMore
 Order printed copiesOrder
Views269
Downloads710
Citations

Related articles

No related content is available yet for this article.

Article of the Year Award: Outstanding research contributions of 2021, as selected by our Chief Editors. Read the winning articles.