A Novel Method (CiMON) for Continuous Intra-Abdominal Pressure Monitoring: Pilot Test in a Pig Model

Table 2

Statistics of comparison between different device readings in the 25° head-up elevation position.

DeviceBias (mmHg)Precision (mmHg)Limits of agreement (mmHg)Percentage error

IAPCiM−IAPspie−1.11.1 11.0
IAPCiM−IAPdir−0.52.7 26.3
IAPCiM−IAPivp−1.02.2 19.9

Mean difference between different device readings (bias), precision (standard deviation), percentage error (%), and limits of agreement in the 25° head-up elevation position. CiMON balloon-tipped catheter pressure (IAPCiM), Spiegelberg balloon-tipped catheter pressure (IAPspie), direct intraperitoneal pressure (IAPdir), and intravesical pressure (IAPivp) measurement.