ROC area under the curve (AUC) values for detection of increased and decreased percentage changes of CO relative to FP where is significance of difference between the two measures. An AUC of 1 represents perfect sensitivity, 0.7 represents clinically acceptable sensitivity to change, while 0.5 is a random relationship.

−40%−30%−20%−15%−10%−5%+5%+10%+15%+20%+30%+40%

PAC0.8550.8110.6350.5490.5340.4960.5240.5850.6310.6590.6310.621
USCOM0.8970.8810.8850.8570.8420.7140.7080.7540.8120.8140.7900.897
0.450.30.0040.0000.0000.0000.0000.0050.010.040.0940.06

We are committed to sharing findings related to COVID-19 as quickly as possible. We will be providing unlimited waivers of publication charges for accepted research articles as well as case reports and case series related to COVID-19. Review articles are excluded from this waiver policy. Sign up here as a reviewer to help fast-track new submissions.