Canadian Journal of Gastroenterology
Volume 22 (2008), Issue 4, Pages 365-368
Original Article

High-Resolution Esophageal Manometry: A Time Motion Study

Daniel C Sadowski1 and Linda Broenink2

1Division of Gastroenterology, Royal Alexandra Hospital, Canada
2Gastrointestinal Motility Laboratory, Walter C MacKenzie Health Sciences Centre, Edmonton, Alberta, Canada

Received 17 October 2007; Accepted 26 November 2007

Copyright © 2008 Hindawi Publishing Corporation. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


INTRODUCTION: High-resolution manometry (HRM) of the esophagus is a new technique that provides a more precise assessment of esophageal motility than conventional techniques. Because HRM measures pressure events along the entire length of the esophagus simultaneously, clinical procedure time should be shorter because less catheter manipulation is required. According to manufacturer advertising, the new HRM system is more accurate and up to 50% faster than conventional methods.

OBJECTIVE: To test the hypothesis that clinical testing with HRM requires less procedural time than a standard water perfusion (WP) method.

METHODS: Forty-one consecutive patients were studied (20 underwent WP and 21 underwent HRM). Using time-motion analysis, the start and end times for each task associated with performing the study were recorded. Patient discomfort and study quality were also assessed by using five- and four-point qualitative scales, respectively.

RESULTS: Total procedure time was reduced on average by 25.6% in the HRM group (from 41.8 minutes with WP to 30.7 minutes with HRM, P<0.05). There was no significant difference in the discomfort scores reported by the study subjects and no difference in study quality.

CONCLUSIONS: HRM requires less time to complete than conventional manometry and should therefore shorten the wait-times of patients scheduled for esophageal manometry and have a significant impact on the cost of performing this commonly used clinical investigation.