Research Article

Investigation of Attenuation Correction for Small-Animal Single Photon Emission Computed Tomography

Table 1

Assessing the results after corrections of rat-sized phantom with five different parameters.

MeanCVRMSENRMSEMPE

Original0.536120.037700.066750.11129
SC only0.378460.052510.222170.37045
Chang + SC0.581650.041940.030350.05060
CTAC + SC0.603980.042570.026030.043400.00706
CTIACR + SC0.607570.027890.018650.031100.01305
CTMAC + SC0.622590.037550.032670.054480.03810

The true value of the activity concentration is 0.5997 MBq/mL.