Research Article

Evaluation of a Model-Based Hemodynamic Monitoring Method in a Porcine Study of Septic Shock

Table 2

The median and range of the measured left and right end diastolic volumes (LVEDV, RVEDV) and maximum left and right ventricular pressures ( lv,max, rv,max). Bias and precision metrics (2SD), median percentage errors with 5th and 95th percentile bounds, and correlation coefficients of the modelled LVEDV, RVEDV, lv,max, and rv,max.

Median (range)Bias ± 2SD % Error (5th–95th percentile)

LVEDV97.7 mL (55.0–141.2)−3.4 ± 10.2 mL3.5% (0.5–10.9) 0.98
RVEDV93.3 mL (44.8–138.2)3.8 ± 10.2 mL4.9% (0.4–16.3) 0.96
86.1 mm Hg (63.2–121.8)−1.6 ± 12.8 mm Hg3.8% (0.4–16.4)0.96
38.7 mm Hg (18.0–60.7)4.5 ± 12.9 mm Hg14.8% (2.6–29.9)0.76