Computational and Mathematical Methods in Medicine / 2015 / Article / Tab 1 / Research Article
An Adaptive Thresholding Method for BTV Estimation Incorporating PET Reconstruction Parameters: A Multicenter Study of the Robustness and the Reliability Table 1 PET/CT scanners: main technical characteristics and physical performances.
Discovery ST (general electric) S1-2 Discovery STE (general electric) S3 Discovery 600 (general electric) S4-5 Discovery 690 (general electric) S6 Biograph16 Hi-REZ (Siemens) S7 Biograph 6 True V (Siemens) S8 Gemini XL (Philips) S9 Gemini TF (Philips) S10 Detector ring diameter (cm) 88.6 88.6 80.1 80.1 83.0 83.0 88.5 90.3 Detector material BGO BGO BGO LYSO LSO LSO GSO LYSO Acquisition mode 2D/3D 2D/3D 3D 3D 3D 3D 3D 3D Number of individual crystals 10.080 13.440 12.288 13824 24.336 32448 17.864 28.336 Number of crystal/ring 420 560 512 576 624 624 616 NA Number of image planes 47 47 47 47 81 109 90 90 Crystal size (mm3 ) Patient port diameter (cm) 70 70 70 70 70 70 70 71.7 Axial field of view (cm) 15.7 15.7 15.7 15.7 16.2 21.8 18.0 18.0 Transaxial filed view (cm) 70 70 70 70 58.5 60.4 57.6 57.6 Axial sampling interval (mm) 3.27 3.27 3.27 3.27 2.0 2.0 2.0 2.0 Coincidence window width (ns) 11.7 9.3 9.0 4.9 4.5 4.5 7.5 6.0 Lower energy threshold (keV) 375 425 425 425 425 425 410 440 Physical performances Transverse resolution FWHM (mm) at 1 cm 6.29 5.1 4.9 4.70 4.61 4.1 5.2 4.8 FWHM (mm) at 10 cm 6.82 5.7 5.6 5.06 5.34 4.8 5.8 5.0 Axial resolution FWHM (mm) at 1 cm 5.68 5.2 5.6 4.74 5.10 4.7 5.8 4.8 FWHM (mm) at 10 cm 6.05 5.9 6.4 5.55 5.93 5.7 6.6 5.2 System sensitivity (cps/KBq) 8.99 8.8 9.6 7.5 4.87 8.0 8.0 6.6 Scatter fraction (%) 45 34 36.6 37 34.1 32.7 35 27