An Adaptive Thresholding Method for BTV Estimation Incorporating PET Reconstruction Parameters: A Multicenter Study of the Robustness and the Reliability

Table 1

PET/CT scanners: main technical characteristics and physical performances.

Discovery ST
(general electric)
S1-2
Discovery STE
(general electric)
S3
Discovery 600
(general electric)
S4-5
Discovery 690
(general electric)
S6
Biograph16 Hi-REZ
(Siemens)
S7
Biograph 6 True V
(Siemens)
S8
Gemini XL
(Philips)
S9
Gemini TF
(Philips)
S10

Detector ring diameter (cm)88.688.680.180.183.083.088.590.3
Detector materialBGOBGOBGOLYSOLSOLSOGSOLYSO
Acquisition mode2D/3D2D/3D3D3D3D3D3D3D
Number of individual crystals10.08013.44012.2881382424.3363244817.86428.336
Number of crystal/ring420560512576624624616NA
Number of image planes47474747811099090
Crystal size (mm3)
Patient port diameter (cm)7070707070707071.7
Axial field of view (cm)15.715.715.715.716.221.818.018.0
Transaxial filed view (cm)7070707058.560.457.657.6
Axial sampling interval (mm)3.273.273.273.272.02.02.02.0
Coincidence window width (ns)11.79.39.04.94.54.57.56.0
Lower energy threshold (keV)375425425425425425410440
Physical performances
Transverse resolution
 FWHM (mm) at 1 cm6.295.14.94.704.614.15.24.8
 FWHM (mm) at 10 cm6.825.75.65.065.344.85.85.0
Axial resolution
 FWHM (mm) at 1 cm5.685.25.64.745.104.75.84.8
 FWHM (mm) at 10 cm6.055.96.45.555.935.76.65.2
System sensitivity (cps/KBq)8.998.89.67.54.878.08.06.6
Scatter fraction (%)453436.63734.132.73527