Research Article

Phantom Validation of Tc-99m Absolute Quantification in a SPECT/CT Commercial Device

Table 2

SPECT acquisitions of the NEMA/IEC NU2 phantom: Activity concentration in the background () and spherical inserts (), total activity deviation (), calibration factor (Bg. cal) and coefficient of variation (COV), lung variability ( lung), and hot contrast () for the 22 mm diameter hot insert.

NEMA
scan number
(%)Bg. calCOV% lung% (%)
(22 mm)

Scan 123.27196.872.51.06205.7865
Scan 218.30154.821.11.03245.7067
Scan 312.32104.20−10.98287.7965
Scan 41.4612.3641.015520.4765