Research Article

Prediction-Based Compensation for Gate On/Off Latency during Respiratory-Gated Radiotherapy

Table 1

Gate on and off latencies of some gating systems.

MonitorLinacGate on delayGate off delayReference

Abches (APEX)Elekta Synergy336 ms88 ms[4]
AlignRT (VisionRT)Varian Clinac iX356 ms529 ms[5]
Calypso (Varian)Varian Clinac iX209 ms60 ms[5]
Catalyst (C-RAD)Elekta Synergy851 ms215 ms[6]