The Effect of CT Scan Parameters on the Measurement of CT Radiomic Features: A Lung Nodule Phantom Study
Table 4
Regression coefficients in the comparison between different slice thickness, mAs, and reconstruction algorithm in −630 HU nodule phantoms.
| Image features | Constant | Slice thickness | mAs | Reconstruction algorithm | 5.00 (ref.) | 2.50 | 1.25 | 30 (ref.) | 120 | Lung (ref.) | Standard |
| Histogram | | Mean | 0.617 | 0 | 0.247† | 0.277† | 0 | −0.011 | 0 | −0.154† | | Stddev | 0.369 | 0 | 0.036 | 0.158† | 0 | −0.149† | 0 | −0.279† | | Variance | 0.230 | 0 | 0.034 | 0.154† | 0 | −0.125† | 0 | −0.200† | | Skewness | −0.023 | 0 | 0.209† | 0.256† | 0 | 0.143† | 0 | 0.224† | | Kurtosis | 0.985 | 0 | −0.146† | −0.098† | 0 | −0.149† | 0 | −0.436† | | Energy | 0.061 | 0 | 0.063† | −0.021 | 0 | 0.189† | 0 | 0.361† | | Entropy | 0.664 | 0 | 0.014 | 0.122† | 0 | −0.196† | 0 | −0.394† |
| GLCM | | Contrast | 0.237 | 0 | 0.111† | 0.238† | 0 | −0.108† | 0 | −0.206† | | Dissimilarity | 0.349 | 0 | 0.121† | 0.251† | 0 | −0.141† | 0 | −0.292† | | Homogeneity | 0.193 | 0 | −0.039† | −0.154† | 0 | 0.191† | 0 | 0.422† | | ASM | 0.027 | 0 | −0.004 | −0.060† | 0 | 0.109† | 0 | 0.166† | | Energy | 0.080 | 0 | −0.015 | −0.086† | 0 | 0.139† | 0 | 0.233† | | Probability max | 0.052 | 0 | −0.008 | −0.036† | 0 | 0.084† | 0 | 0.122† | | Entropy | 0.788 | 0 | 0.042 | 0.130† | 0 | −0.153† | 0 | −0.284† | | | Correlation | 0.752 | 0 | −0.100† | −0.228† | 0 | 0.109† | 0 | 0.204† |
| GLRLM | | LRE | 0.068 | 0 | 0.014 | −0.054† | 0 | 0.130† | 0 | 0.258† | | GLN | 0.033 | 0 | 0.033† | −0.018 | 0 | 0.125† | 0 | 0.223† | | RLN | 0.891 | 0 | −0.004 | 0.078† | 0 | −0.163† | 0 | −0.318† | | LGRE | 0.167 | 0 | −0.106† | −0.086† | 0 | −0.018 | 0 | 0.011 | | HGRE | 0.639 | 0 | 0.227† | 0.269† | 0 | −0.030† | 0 | −0.144† |
|
|
|