Determination of Glucosamine in Human Plasma by High-Performance Liquid Chromatography-Atmospheric Pressure Chemical Ionization Source-Tandem Mass Spectrometry

Table 2

Precision and accuracy for the analysis of glucosamine in human plasma.

106.5 ng/mL852.3 ng/mL3409 ng/mL
Determined mean concentration
(ng/mL)±S.D.
R.S.D. (%)Mean accuracy (%)Determined mean concentration
(ng/mL)±S.D.
R.S.D. (%)Mean accuracy (%)Determined mean concentration
(ng/mL)±S.D.
R.S.D. (%)Mean accuracy (%)

Day 1 ( 𝑛 = 5 ) 1 0 5 . 3 ± 7 . 4 7.0598.9 8 9 1 . 7 ± 1 0 3 . 4 11.699.5 3 3 4 3 ± 1 2 3 3.6793.6
Day 2 ( 𝑛 = 5 ) 1 0 6 . 0 ± 6 . 1 5.80104.6 8 9 2 . 3 ± 2 8 . 2 3.16104.7 3 1 4 2 ± 1 0 1 3.2498.5
Day 3 ( 𝑛 = 5 ) 9 9 . 7 ± 7 . 5 7.5698.1 8 3 9 . 2 ± 2 9 . 8 3.5592.2 3 0 9 9 ± 3 0 1 9.7190.9
Interday ( 𝑛 = 1 5 ) 1 0 3 . 7 ± 8 . 4 8.1297.4 8 7 4 . 4 ± 4 3 . 2 4.94102.6 3 1 9 5 ± 2 4 4 7.6593.7