Table 5

Validation results of developed and optimized reverse phase chromatographic method.

ParameterPYH (avg. %)DOX (avg. %)

SpecificityNo interference

Accuracy
50%102.23100.10
75%100.6098.47
100%98.72100.61
125%100.5099.87
150%100.61100.37

Precision—Repeatability
0 hrs99.2699.34
8 hrs99.1899.34
16 hrs99.2099.33

Precision—Intermediate
Day-199.29100.42
Day-299.4299.89
Day-398.9098.37
Instrument-199.4199.34
Instrument-2101.24100.25
Analyst-1100.2099.35
Analyst-2100.2699.34
Lab-199.4299.31
Lab-299.9699.33

Robustness
Flow rate0.9 mL/min99.2899.69
1.1 mL/min99.25100.24
Wavelength+2 nm98.2099.31
−2 nm101.6899.40
Mobile phase+2%99.73100.11
−2%100.21100.26
pH+0.05100.2399.96
−0.0599.42100.28
Assay100.24101.39