Development and Validation of New RP-HPLC Method for Simultaneous Determination of Methyl and Propyl Parabens with Levetiracetam in Pure Form and Pharmaceutical Formulation
Table 1
Regression and statistical parameters for determination of LTC, MHB, and PHB using the proposed HPLC method.
Parameter
Levetiracetam
Methyl paraben
Propyl paraben
System suitability
± SD (min)
1.561 ± 8.22 × 10−3
5.855 ± 0.6 × 10−2
7.844 ± 7.8 × 10−2
USP plate count ()
3.36 × 103
3.59 × 104
9.04 × 104
Resolution (Rs)
—
3.7
4.45
Symmetry factor ()
1.08
1.04
0.996
Selectivity (α)
—
4.91
1.08
Linearity and regression data
Linearity range (µgmL−1)
500–1500
10–30
1–3
LOD (µgmL−1)
0.0078
0.025
0.029
LOQ (µgmL−1)
0.023
0.076
0.088
Slope ()
220896
21738
31501
Intercept ()
3880
−3407
5574.4
Correlation coefficient ()
0.9996
0.9999
0.9994