Fumiaki Nakao, Masashi Kanemoto, Jutaro Yamada, Kazuhiro Suzuki, Hidetoshi Tsuboi, Takashi Fujii, "A Rare Case of Complete Stent Fracture, Coronary Arterial Transection, and Pseudoaneurysm Formation Induced by Repeated Stenting", Case Reports in Cardiology, vol. 2015, Article ID 192853, 4 pages, 2015. https://doi.org/10.1155/2015/192853

A Rare Case of Complete Stent Fracture, Coronary Arterial Transection, and Pseudoaneurysm Formation Induced by Repeated Stenting

Academic Editor: Mohammad R. Movahed
Received29 Jun 2015
Revised12 Sep 2015
Accepted28 Sep 2015
Published12 Oct 2015

Abstract

This report describes a rare asymptomatic case of complete stent fracture, coronary arterial transection, and pseudoaneurysm formation in response to repeated stenting. The proximal and distal ends of transected coronary artery were closed, and distal bypass was performed. Coronary arterial transection can occur in patients with repeated stenting as a long-term adverse event.

1. Introduction

Stent fracture after drug-eluting stent (DES) deployment is an important issue, because it is strongly associated with restenosis, target legion revascularization, and stent thrombosis [1]. A report of autopsy cases with DES deployment showed stent fracture in 29% of lesions and restenosis or stent thrombosis in 67% of cases with gapped stent fracture [2].

Stent fracture can also lead to coronary pseudoaneurysm formation, which can be life-threatening [3]. The incidence of coronary pseudoaneurysm formation after DES deployment is 0.3–4.5% [4]. Management strategies for coronary pseudoaneurysm include observation, surgical treatment and interventional treatment, such as coil embolization and deployment of a polytetrafluoroethylene- (PTFE-) covered stent [35].

2. Case Report

A 61-year-old male undergoing chronic hemodialysis had previously underwent rotational atherectomy and stenting (TAXUS Liberte, Boston Scientific Co.) for a long, severely calcified lesion of the right coronary artery (RCA) (first percutaneous coronary intervention [PCI#1], Figure 1). Six months later, the patient underwent emergent restenting (Cypher, Cordis) for probable stent thrombosis of the mid-RCA with ST elevation (second PCI [PCI#2], Figure 2). Four months later, he underwent emergent repeat stenting (Xience V, Abbott Vascular) for probable stent thrombosis of the mid-RCA with ST elevation (third PCI [PCI#3], Figure 3). Two months later, he was admitted for follow-up coronary angiography (CAG) and was noted to be asymptomatic. CAG showed pseudoaneurysm formation in the mid-RCA (see Figures 4(a), 4(b), and 4(c) and see Clip  1 in Supplementary Material available online at http://dx.doi.org/10.1155/2015/192853), and X-ray fluorography showed complete stent fracture (Figure 4(d)). Coronary transection was suspected, because of findings of complete stent fracture and contrast media oozing all around the part of stent fracture.

3. Discussion

Risk factors for stent fracture include RCA stenting, long stenting, overlapped stenting, and stenting on a hinge point [6, 7]. The present patient underwent long and overlapped stenting within the RCA and therefore was at high risk for stent fracture. Drugs and polymers of DES may induce vascular inflammation and delay vascular healing [8], and they also can contribute to pseudoaneurysm formation. In this case, the vessel wall was likely exposed to a relatively high dose of DES drug and polymer (due to three overlapping stents).

Surgical treatment and a PTFE-covered stent deployment were considered for this case. However, a guidewire could perforate the wall of the pseudoaneurysm, and deployment of the PTFE-covered stent might be difficult, because previous procedures required the mother and child (4 in 6) technique. If repeated stenting for stent fracture was performed, stent fracture might occur repeatedly, leading to lethal stent thrombosis or blow-out rupture of the pseudoaneurysm. Therefore, surgical management was selected for this case. During surgery, the pseudoaneurysm was visualized in the visceral adipose tissue (arrow heads, Figure 5(a)). After the pseudoaneurysm was opened (Figure 5(b)), coronary transection was confirmed (arrows, Figure 5(b)). The proximal and distal transected ends of the mid-RCA could not be ligated because of protrusion of the overlapped fractured struts (arrows, Figure 5(c)). Therefore, proximal and distal transected ends of the mid-RCA were closed (Figure 5(d)), and distal bypass was performed.

In conclusion, this case report described a rare asymptomatic case of complete stent fracture, coronary arterial transection, and pseudoaneurysm formation in response to repeated stenting. Coronary arterial transection can occur in patients with repeated stenting as a long-term adverse event.

Conflict of Interests

The authors declare that there is no conflict of interests with regard to this report.

Acknowledgments

The authors thank Tooru Ueda, Takamasa Oda, and Yasuhiro Ikeda of the Department of Cardiology, Ymaguchi Grand Medical Center, for their support and thank Takayuki Okamura and Masafumi Yano of the Division of Cardiology, Department of Medicine and Clinical Science, Yamaguchi University Graduate School of Medicine, for their helpful advice.

Supplementary Materials

Follow-up coronary angiography of right coronary artery shows pseudoaneurysm formation and complete stent fracture. Coronary transection is suspected, because of findings of complete stent fracture and contrast media oozing all around the part of stent fracture.

  1. Supplementary Material

References

  1. S. E. Lee, M. H. Jeong, I. S. Kim et al., “Clinical outcomes and optimal treatment for stent fracture after drug-eluting stent implantation,” Journal of Cardiology, vol. 53, no. 3, pp. 422–428, 2009. View at: Publisher Site | Google Scholar
  2. G. Nakazawa, A. V. Finn, M. Vorpahl et al., “Incidence and predictors of drug-eluting stent fracture in human coronary artery a pathologic analysis,” Journal of the American College of Cardiology, vol. 54, no. 21, pp. 1924–1931, 2009. View at: Publisher Site | Google Scholar
  3. Y. Kawai, M. Kitayama, H. Akao, A. Motoyama, T. Tsuchiya, and K. Kajinami, “A case of coronary rupture and pseudoaneurysm formation after fracture of implanted paclitaxel-eluting stents,” Cardiovascular Intervention and Therapeutics, pp. 1–7, 2015. View at: Publisher Site | Google Scholar
  4. S. Bajaj, R. Parikh, A. Hamdan, and M. Bikkina, “Covered-stent treatment of coronary aneurysm after drug-eluting stent placement,” Texas Heart Institute Journal, vol. 37, no. 4, pp. 449–454, 2010. View at: Google Scholar
  5. A. Maroo, P. A. Rasmussen, T. J. Masaryk, S. G. Ellis, A. M. Lincoff, and S. Kapadia, “Stent-assisted detachable coil embolization of pseudoaneurysms in the coronary circulation,” Catheterization and Cardiovascular Interventions, vol. 68, no. 3, pp. 409–415, 2006. View at: Publisher Site | Google Scholar
  6. H. Doi, A. Maehara, G. S. Mintz et al., “Classification and potential mechanisms of intravascular ultrasound patterns of stent fracture,” The American Journal of Cardiology, vol. 103, no. 6, pp. 818–823, 2009. View at: Publisher Site | Google Scholar
  7. S. Kuramitsu, M. Iwabuchi, T. Haraguchi et al., “Incidence and clinical impact of stent fracture after everolimus-eluting stent implantation,” Circulation: Cardiovascular Interventions, vol. 5, no. 5, pp. 663–671, 2012. View at: Publisher Site | Google Scholar
  8. R. Virmani, F. Liistro, G. Stankovic et al., “Mechanism of late in-stent restenosis after implantation of a paclitaxel derivate-eluting polymer stent system in humans,” Circulation, vol. 106, no. 21, pp. 2649–2651, 2002. View at: Publisher Site | Google Scholar

Copyright © 2015 Fumiaki Nakao et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


More related articles

690 Views | 429 Downloads | 2 Citations
 PDF Download Citation Citation
 Download other formatsMore
 Order printed copiesOrder

Related articles

We are committed to sharing findings related to COVID-19 as quickly as possible. We will be providing unlimited waivers of publication charges for accepted research articles as well as case reports and case series related to COVID-19. Review articles are excluded from this waiver policy. Sign up here as a reviewer to help fast-track new submissions.