Case Report

Subacute Right Ventricular Perforation by Pacemaker Lead Causing Left-Sided Hemothorax and Epicardial Hematoma

Table 1


Normal valuesExpected change in perforationWhy the parameter may change differently even in perforation

Impedance400–1000 OhmsUsually decreasedMay increase if the lead ends in an air-filled space
Sensing (R-wave amplitude)At least 5 mVUsually decreasedMay increase if the lead becomes parallel to the incoming electric current vector
Capture thresholdLess than 1 volt at a pulse width of 0.5 millisecondsUsually there is loss of captureMay remain the same if the lead has not moved a long distance from the heart