Radiation Dose Optimization in Interventional Cardiology: A Teaching Hospital Experience
Table 5
Quantitative image quality results using the CIRS model 901 20 cm phantom for standard NEMA measurements.
| | Phase 1 | Phase 2 | Phase 3 | Phase 4 | Phase 5 | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions | Fluoroscopy | Acquisitions |
| X-ray parameters | kVp | 81 | 70 | 78 | 70 | 76 | 70 | 76 | 70 | 76 | 71 | Tube current (mA) | 6.5 | 342 | 5 | 342 | 3 | 342 | 3 | 373 | 3 | 400 | Dose setting | Normal | Normal | Low | Normal | Low | Normal | Low | Normal | Low | Low | Frame rate | 7.5 | 15 | 7.5 | 15 | 5 | 15 | 5 | 10 | 5 | 10 | Added filteration | Al 2.0 + Cu 0.1 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 2.0 + Cu 0.1 |
| Static image quality | Line pairs | 1.8 | 2.2 | 1.8 | 2.2 | 1.4 | 2.2 | 1.4 | 2 | 1.4 | 2 | Iodine group 1 | 6 | 8 | 6 | 8 | 6 | 8 | 6 | 6 | 6 | 8 | Iodine group 2 | 4 | 6 | 0 | 6 | 0 | 6 | 0 | 6 | 0 | 6 | Iodine group 3 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | Iodine group 4 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | Air cylinders | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | Aluminum cylinders | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 3 | 4 | 3 |
| X-ray parameters | kVp | 75 | 71 | 77 | 71 | 77 | 71 | 77 | 72 | 77 | 73 | Tube current (mA) | 5 | 432 | 5.7 | 432 | 3 | 432 | 3 | 400 | 3 | 399 | Dose setting | Normal | Normal | Low | Normal | Low | Normal | Low | Normal | Low | Low | Frame rate | 7.5 | 15 | 7.5 | 15 | 5 | 15 | 5 | 10 | 5 | 10 | Added filteration | Al 2.0 + Cu 0.1 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 1.0 + Au 0.01 | Al 1.5 + Cu 0.6 | Al 2.0 + Cu 0.1 |
| Temporal image quality | Moving wires | 3 | 4 | 3 | 4 | 3 | 4 | 3 | 4 | 3 | 4 |
|
|