Radiation Dose Optimization in Interventional Cardiology: A Teaching Hospital Experience

Table 5

Quantitative image quality results using the CIRS model 901 20 cm phantom for standard NEMA measurements.

Phase 1Phase 2Phase 3Phase 4Phase 5
FluoroscopyAcquisitionsFluoroscopyAcquisitionsFluoroscopyAcquisitionsFluoroscopyAcquisitionsFluoroscopyAcquisitions

X-ray parameterskVp81707870767076707671
Tube current (mA)6.53425342334233733400
Dose settingNormalNormalLowNormalLowNormalLowNormalLowLow
Frame rate7.5157.515515510510
Added filterationAl 2.0 + Cu 0.1Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 2.0 + Cu 0.1

Static image qualityLine pairs1.82.21.82.21.42.21.421.42
Iodine group 16868686668
Iodine group 24606060606
Iodine group 30000000000
Iodine group 40000000000
Air cylinders4444444444
Aluminum cylinders4444444343

X-ray parameterskVp75717771777177727773
Tube current (mA)54325.7432343234003399
Dose settingNormalNormalLowNormalLowNormalLowNormalLowLow
Frame rate7.5157.515515510510
Added filterationAl 2.0 + Cu 0.1Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 1.0 + Au 0.01Al 1.5 + Cu 0.6Al 2.0 + Cu 0.1

Temporal image qualityMoving wires3434343434