Development and Validation of an ELISA at Acidic pH for the Quantitative Determination of IL-13 in Human Plasma and Serum

Table 6

Stability assessment of endogenous IL-13 in human plasma. Two different human plasma samples (3.75 and 1.16 pg/mL of endogenous IL-13) were stored undiluted under different temperatures and time periods or subjected to freeze and thaw cycles. Recovery of IL-13 concentration was calculated using the concentration measured at Day 1 (T0) as a reference value.

Storage conditions Mean recovery (%)Range of recovery (%)

Short-term stability25C, 1 week77.272.3–81.9
4C, 1 week82.374.0–94.3
−20C, 1 month80.777.4–83.3
−80C, 1 month76.173.9–78.2
Freeze/thaw stability3 cycles at −20C92.273.6–112.5
3 cycles at −80C96.684.3–105.5