[Retracted] System Accuracy Evaluation of the GlucoRx Nexus Voice TD-4280 Blood Glucose Monitoring System

Table 3

Overall determination of the accuracy of the GlucoRx Nexus TD-4280 blood glucose monitoring device relative to the accuracy limits defined in both the ISO 15197:2003 and ISO 15197:2013 guidelines.

DIN EN ISO 15197:2003DIN EN ISO 15197:2013

Accuracy limits set by guideline±15 mg/dL (0.83 mmol/L) and ±20%±15 mg/dL (0.83 mmol/L) and ±15%
Percentage of results required to be within this limit for meter to be deemed accurate≥95%≥95%
Number of results obtained within this limit97/10591/105
Percentage of results obtained within this limit (%)92.486.7
Are ≥95% of the readings within the accuracy limits set in guidelines? (Yes/no)NoNo