Research Article
Validation of an LC-MS/MS Method for Urinary Lactulose and Mannitol Quantification: Results in Patients with Irritable Bowel Syndrome
Table 2
Between-run and within-run accuracy and precision.
| Analyte | Nominal concentration (mg/L) | Imprecision | Accuracy | Intra-assay CV (%) | Interassay CV (%) | Intra-assay (%) | Interassay (%) |
| Mannitol | 100 | 5.1 | 12.1 | 96.4 | 95.9 | 400 | 4.9 | 11.5 | 93.2 | 94.7 | 600 | 3.8 | 12.4 | 96.8 | 96.7 |
| Lactulose | 50 | 2.9 | 5.1 | 85.6 | 85.8 | 100 | 4.5 | 5.6 | 91.3 | 90.9 | 200 | 4.2 | 6.3 | 94.3 | 94.3 |
|
|