The Use of Catheter Mount Will Result in More Reliable Carbon Dioxide Monitoring under Fluid Exposing Conditions

Table 2

ETCO2 by using pDCCM and dDCCM.

CapnometerWater spraysETCO2 (mmHg)
†pDCCM‡dDCCMMean difference§p-value
(n=25)(n=25)

Mainstream0 ml34.5 ± 6.531.9 ± 4.9-2.5 ± 5.20.09
5 ml19.0 ± 23.529.5 ± 7.010.4 ± 24.1<0.001
10 ml21.2 ± 24.533.8 ± 14.812.5 ± 28.4<0.001

Microstream0 ml32.5 ± 3.732.7 ± 4.30.2 ± 2.50.83
5 ml13.9 ± 15.230.5 ± 5.116.6 ± 16.7<0.001
10 ml11.4 ± 14.429.9 ± 4.318.5 ± 16.0<0.001

Values are given as mean ± SD.
†pDCCM, proximal capnometer of direct connect catheter mount
‡dDCCM, distal capnometer of direct connect catheter mount
§Calculated by repeated measures ANOVA.
Abbreviations: dDCCM, distal capnometer of direct connect catheter mount; ETCO2, End-tidal carbon dioxide; pDCCM, proximal capnometer of direct connect catheter mount; SD, standard deviation