Research Article
Fast MR Imaging of the Paediatric Abdomen with CAIPIRINHA-Accelerated T1w 3D FLASH and with High-Resolution T2w HASTE: A Study on Image Quality
Table 1
Typical MRI scan parameters, as used in this study.
| | Standard 2D FLASH | CAIPIRINHA 3D FLASH | T1w TSE triggered | T2w TIRM | T2w HASTE FS |
| TR, ms | 86.0 | 6.1 | 2000 | 8290 | 1000 | TE, ms | 3.8 | 3.0 | 2.5 | 84 | 102 | Flip angle, ° | 90 | 10 | 15 | 180 | 160 | Scan orientation | Transverse | Transverse | Transverse | Coronal | Coronal | In-plane resolution, mm | 1.3 × 1.3 | 1.0 × 1.0 | 0.7 × 0.7 | 0.8 × 0.8 | 1.0 × 1.0 | Slice thickness, mm | 5 | 3–5 | 4 | 4 | 4 | FOV, mm | 250 | 250 | 350 | 350 | 350 | iPAT | None | Acceleration factor = 4 | None | Acceleration factor = 2 | Acceleration factor = 2 | Breath-hold | 22 s | 15 s | Respiratory-triggered | 20 s | 21 s | Scan time | 4 min 3 s | 15 s | 6 min 10 s | 3 min 5 s | 42 s |
|
|
TR: repetition time, TE: echo time, FOV: field of view, and iPAT: Integrated Parallel Acquisition Techniques.
|