Clinical Study

Third-Generation Capsule Endoscopy Outperforms Second-Generation Based on the Detectability of Esophageal Varices

Table 3

Accuracy of capsule endoscopy for the diagnosis of esophageal varices.
(a)

SB2 findingsEGD findingsTotal
+

+36036
194463

Total554499

value, 0.63; sensitivity, 65%; specificity, 100%; positive predictive value, 100%; negative predictive value, 70%.
(b)

SB3 findingsEGD findingsTotal
+

+47047
111829

Total581876

value, 0.67; sensitivity, 81%; specificity, 100%; positive predictive value, 100%; negative predictive value, 62%.
EGD: esophagogastroduodenoscopy; SB2: second-generation PillCam SB2; SB3: third-generation PillCam SB3.