Research Article

Quantitation by Portable Gas Chromatography: Mass Spectrometry of VOCs Associated with Vapor Intrusion

Table 4

Method validation results.

CompoundMass (m/z) (s) % RSDLinear Range (ppbv)Avg. Response FactorMDL
ppbvmg m−3ng

Benzene78410.9995.90–501 3.2100.25
Carbon tetrachloride117420.9992.50–499 2.7110.28
(E/Z)-1,2-Dichloroethylene96360.98711.30–499 12.147.91.19
Ethyl chloride66340.99017.30–250 16.543.61.08
Methylene chloride84350.98514.10–251 27.696.02.38
Trichloroethylene97450.9986.80–500 2.1110.27
Tetrachloroethylene166630.9990.80–501 0.740.1
Toluene92540.9986.80–502 1.86.90.17
Vinyl chloride62340.9993.30–250 13330.81
o, m, p-Xylenes10682 0.9995.30–500 3.2130.34

Retention time. Sample concentration time of 15 s at a 99 cm3 min−1 flow rate. A of 82 s with a window wide enough to include all isomeric xylenes.