Interday precision over four days (% relative standard deviation (% RSD)) and accuracy (% error) of the LC-MS/MS assay for quantification of lansoprazole (n = 20 for each concentration).

Concentration of lansoprazole added (μg/mL)Concentration of lansoprazole found (μg/mL)% Relative standard deviation% Error

 55.07 + 0.183.551.18
 1010.13 + 0.201.991.34
 1514.84 + 0.583.931.06
 2020.56 + 0.462.232.81
 2524.44 + 0.893.632.23