Ultra High-Mass Resolution Paper Spray by Fourier Transform Ion Cyclotron Resonance Mass Spectrometry

Table 1

PEG standards identified using paper spray mass spectrometry. The PEG mass spectrum was obtained with both an excitation and detection frequency range of 1.5 MHz.

PEGExperimental m/z Theoretical m/z Mass measurement error (Δppm)Mass resolving power (M/ΔM50%)[H(OCH2CH2)nOH+Na]+

1009.5776391009.57651−0.1246309
1053.602741053.602730.0142870
1097.628861097.62894−0.0740306
1141.655271141.655150.0238791
1185.681281185.68137−0.0837199
1229.707791229.707580.1735953
1273.733911273.73380.0934776
1317.760351317.760010.2633645
1361.786401361.786230.1232579
1405.812321405.81244−0.0931288
1449.838681449.838660.0130101
1493.865391493.864870.3529219
1537.891721537.891090.4128219
1581.917421581.91730.0827254
1625.943581625.943520.0426391
1669.969681669.96973−0.0325567
1713.995201713.99595−0.4424691