Research Article

Estimation of Measurement Uncertainties for the DGT Passive Sampler Used for Determination of Copper in Water

Table 3

Uncertainty budget for showing relative uncertainties for the variables and the combined standard uncertainty.

SymbolSource of uncertaintyType*Standard uncertainty DistributionDivisorRelative uncertainty

Estimated mass from ICP-MS analysisA gNormal10.008
Recovery factorB0.0293 Rectangular0.064
Uc (M)Combined standard uncertaintyA gNormal0.038

Note: type of uncertainty refers to types A and B, using standard vocabulary for statistically evaluated uncertainty (A) and uncertainty evaluated by other methods (B).