Research Article

Development and Validation of Chronopotentiometric Method for Imidacloprid Determination in Pesticide Formulations and River Water Samples

Table 1

Linear ranges for chronopotentiometric determination of imidacloprid using thin film mercury and glassy carbon electrode.

ElectrodeConcentration range [mg/L]Reduction current [A]SlopeIntercept

Thin film Mercury Electrode0.8–2.0−3.80.2570.8980.01870.01310.9989
2.0–10.0−6.20.1320.3900.00200.01950.9975
10.0–30.0−16.80.086−0.2800.00220.01230.9980

Glassy Carbon Electrode7.0–15.0−6.60.0460.3290.00310.01290.9976
10.0–40.0−14.00.0320.0380.00070.00120.9987
40.0–70.0−18.40.051−0.7300.00020.02200.9986

represents the standard deviation of the slope in s⋅L/mg, .
represents the standard deviation of the intercept in s, .