Research Article

Rapid Determination of Isomeric Benzoylpaeoniflorin and Benzoylalbiflorin in Rat Plasma by LC-MS/MS Method

Table 3

Precision, accuracy, recovery, and matrix effect of BP and BA in QC samples ().

CompoundConcentration/ng mL−1Intraday precision and accuracyInterday precision and accuracyRecovery, %Matrix effect, %
Measured concentration/ng mL−1RSD, %RE, %Measured concentration/ng mL−1RSD, %RE, %

BP10.99 ± 0.1212.480.830.94 ± 0.088.535.5099 ± 6.6298 ± 11.20
22.02 ± 0.126.00−1.252.16 ± 0.073.64−8.0098 ± 8.3199 ± 10.96
500505.16 ± 11.472.27−1.03524 ± 30.285.77−4.8090 ± 10.2094 ± 3.27
750737.66 ± 23.723.211.64750.16 ± 15.272.03−0.0293 ± 14.2998 ± 7.83

BA10.98 ± 0.088.401.660.89 ± 0.033.9610.3397 ± 7.8899 ± 9.23
22.11 ± 0.178.12−5.831.93 ± 0.073.743.1698 ± 5.5296 ± 11.46
500494.83 ± 24.474.941.03494 ± 15.653.161.2097 ± 3.4696 ± 8.58
750753.16 ± 18.252.42−0.42760.66 ± 15.942.09−1.4299 ± 3.9798 ± 11.75