Research Article
LC-MS/MS Quantification of Tramadol and Gabapentin Utilizing Solid Phase Extraction
Table 3
Precision and accuracy results of GBP.
| QC name/nominal concentration | GBP | Intra batch (n=12) | Inter batch (n=24) | % Accuracy | % CV | % Accuracy | % CV |
| LLOQQC/ 10 ng/mL | 96.7 | 6.7 | 98.4 | 3.8 |
| LQC/ 30 ng/ mL | 93.5 | 5.8 | 97.9 | 6.6 |
| MQC/ 2500 ng/mL | 99.1 | 3.2 | 94.1 | 2.9 |
| HQC/ 4500 ng/mL | 95.6 | 2.9 | 100.9 | 5.3 |
|
|