Research Article
Rapid, Validated UPLC-MS/MS Method for Determination of Glibenclamide in Rat Plasma
Table 2
Results of validation studies.
| Validation Parameter | Validation Results | LLOQ | LOQ | MLOQ | HLOQ |
| QC Samples | 10 ng/mL | 20 ng/mL | 160 ng/mL | 1280 ng/mL |
| Accuracy | 117 ± 3.8 % | 109.7 ± 6.7 % | 93.6 ± 0.4 % | 99.3 ± 1.9 % |
| Precision | 4.52 %CV | 5.94 %CV | 1.55 %CV | 2.08 %CV |
| Linearity | 10 – 1280 ng/mL |
| Recovery | 104.2 ± 4.9 % | - - - - - - - - - - - - - - - - - | 100.6 ± 0.9 % | 102.9 ± 5.8 % |
| Stability | 9.9 ± 0.6 ng/mL | 20.3 ± 1.3 ng/mL | 152.5 ± 8.7 ng/mL | 1234 ± 33.0 ng/mL |
|
|