Research Article

Development of RP-HPLC Method for the Simultaneous Quantitation of Levamisole and Albendazole: Application to Assay Validation

Table 1

Optimized chromatographic conditions.

Mobile phaseMixed Phosphate buffer (KH2PO4+K2HPO4): Acetonitrile 30:70
pH3.5
ColumnINERTSIL column, C18 (150x4.6 ID) 5μm
Flow rate1.0 ml/min
Column temperatureRoom temperature (20-25°C)
Sample temperatureRoom temperature (20-25°C)
Wavelength224nm
Injection volume20 μl
Run time10 min
Retention timeAbout 2.350min for Levamisole and 4.055min for Albendazole