Research Article

Development and Validation of an HPLC-MS/MS Method for Rapid Simultaneous Determination of Cefprozil Diastereomers in Human Plasma

Table 2

Assessment of recovery and matrix effect for cis- and trans-cefprozil (n = 6).

CompoundConcentration
(μg/mL)
Recovery (%)
Mean ± SD
IS normalized matrix effect
Mean (%)CV (%)

Cis-cefprozil0.05100.3±5.393.73.0
4.196.2±3.0NANA
12.499.1±7.792.63.8
Trans-cefprozil0.025102.9±4.292.47.7
0.44596.9±4.3NANA
1.33105.1±2.891.66.0

SD: standard deviation; CV: coefficient of variance; NA: not available.