Research Article
Simultaneous Quantification of Methotrexate and Its Metabolite 7-Hydroxy-Methotrexate in Human Plasma for Therapeutic Drug Monitoring
Table 3
Intraday and interday precision and accuracy of MTX and 7-OH-MTX in human plasma (n=5).
| Analyte | Nominal concentration (ng/mL) | Intra-day(n = 5) | Inter-day(n = 15) | Measured concentration (ng/mL) | Precision (RSD%) | Accuracy (RE%) | Measured concentration (ng/mL) | Precision (RSD%) | Accuracy (RE%) |
| MTX | 10 | 9.48 ± 0.23 | 2.47 | -5.22 | 9.30 ± 0.32 | 3.41 | -7.05 | 500 | 514.84 ± 4.56 | 0.89 | 2.97 | 515.67 ± 9.49 | 1.84 | 3.13 | 5000 | 5265.35 ± 92.88 | 1.76 | 5.31 | 5305.42 ± 118.95 | 2.24 | 6.11 | 7-OH-MTX | 10 | 9.69 ± 0.66 | 6.86 | -3.15 | 9.70 ± 0.62 | 6.40 | -3.04 | 500 | 456.43 ± 8.69 | 1.90 | -8.71 | 469.16 ± 14.98 | 3.19 | -6.17 | 5000 | 4772.84 ± 267.84 | 5.61 | -4.54 | 4772.93 ± 189.12 | 3.96 | -4.54 |
|
|