Green Analytical Methods for the Separation of Seven Antihistamines: Application in Separation of Azelastine and Related Impurities in Nasal Solution

Academic Editor: Shaoping Li
Received30 Nov 2018
Accepted16 Jan 2019
Published11 Feb 2019

Abstract

Antihistamines are widely used to alleviate the symptoms caused by allergic reactions. Most of these drugs have zwitteriónicas and/or amphoteric characteristics, which confer additional analytical challenges. This work aimed to develop a single eco-friendly and efficient chromatographic methods for analysis of seven antihistamines, namely, azelastine HCl, desloratadine, ebastine, fexofenadine HCl, ketotifen, loratadine, and olopatadine HCl. The separations were obtained using RP C-18 LUNA (150x4.6mm, 5 μm) column. The mobile phase consisted of acetonitrile and acidified water (pH 2.1) in the following proportion: 15:85, v/v for desloratadine, 25:75, v/v for ketotifen and olopatadine, 32:68, v/v for fexofenadine, 35:65, v/v for azelastine and loratadine, and 45:55, v/v for ebastine. All separations were obtained in less than 7.0 min. A prototype method was fully validated and applied in the assay of azelastine HCl in nasal solutions. The proposed methods for analysis of seven antihistamines are highly efficient, selective, and sensitive. Moreover, all methods can be considered excellent in terms of greenness, with total organic residue < 2.5 mL/analysis. An improved gradient method is also described for separation of azelastine HCl and its related impurities.

1. Introduction

It is estimated that around 400 million people are affected by allergic rhinitis and 300 million by asthma in the world. In Brazil, 30 to 35% of the population presents some type of allergic reaction such as allergic rhinitis, asthma, and atopic dermatitis. Total medical expenses exceed 98 million per year with hospitalizations and are the third cause of hospitalization. These overheads outweigh tuberculosis and AIDS [13].

The second-generation antihistamines are considered as first-line treatment in allergic disorders and do not cause drowsiness and anticholinergic effects. The improved effectiveness, reduced adverse effects, and remarkable drug compliance of modern antihistaminic drugs may be attributed to low lipophilicity and reduced affinity for brain H1 receptors [46].

The second-generation antihistaminic drugs selected in this study are piperidines except for azelastine (methylazepane derivative) and olopatadine (dibenz[b,e]oxepin derivative). The chemical name, molecular structures, and dissociation constants of azelastine HCl (AZE), desloratadine (DESl), ebastine (EB), fexofenadine HCl (FEX), ketotifen (KET), loratadine (LOR), and olopatadine HCl (OLO) are described in Table 1.


AntihistaminesChemical namepKaChemical structure

Azelastine HCl4-[(-chlorophenyl)methyl]-2-(1-methylazepan-4-yl)phthalazin-1-one hydrochloride8.88

Desloratadine8-Chloro-6,11-dihydro-11-(4-piperidinylidene)-5H- benzo[,6]cyclohepta[,2,b]pyridine4.33 and 9.73

Ebastine1-[-(1,1-Dimethylethyl)phenyl]-4-[-(diphenylmethoxy)-1-piperidinyl]-1-butanone8.43

Fexofenidine HCl2-[-[1-hydroxy-4-[-[hydroxy(diphenyl)methyl]-1-piperidyl]butyl]phenyl]-2-methyl-propionic acid hydrochloride4.04 and 9.01

Ketotifen4-(1-Methyl-4-piperidinylidene)-4,9-dihydro-10H-benzo[,5]cyclohepta[1,2-b]thiophen-10-one7.18 and 8.11

LoratadineEthyl 4-(8-chloro-5,6-dihydro-11H-benzo[,6]cyclohepta[,2-b]pyridin-11-ylidene)-1-piperidinecarboxylate4.33

Olopatadine HCl(11Z)-11-[-(Dimethylamino)propylidene]-6,11-dihydrodibenz[b,e]oxepin-2-acetic acid hydrochloride3.78 and 9.76

Several HPLC methods were found in the scientific literature for quantitative determination of second-generation antihistamines, as a single drug (not cited). The British Pharmacopoeia describe monograph on AZE [7], DESL [8], EB [9], FEX [10], KET [11], and LOR [11] while the United States Pharmacopeia contain monograph on AZE [12], DESL [13], LOR [14], and OLO [15]. The cited monographs also describe separation methods for respective organic impurities. Most of these methods are time-consuming, often present peak tailing, make use of buffers salts, and generate a considerable amount of organic waste.

The United States Pharmacopeia describes chromatographic method for separation of azelastine and its related compounds B, D, and E [12]. The separations were obtained on a cyano HPLC column with mobile phase composed of acetonitrile, monobasic potassium phosphate, and octanesulfonic acid sodium salt. The flow rate was 2.0 mL min−1 and detection was made at 210 nm. The system suitability resolution between adjacent peaks is ≥ 1.5 except in case of impurities B and D (≥4.0).

The British Pharmacopoeia [7] and European Pharmacopeia [16] also describe a near identical method for separation of azelastine and its related compounds A, B, C, D, and E.

In the above cases, separations were obtained on a nonconventional column with nitrile groups chemically bonded to porous silica particle with mobile phase composed of buffering and ion pairing salts. Moreover, flow rate is too high, with estimated total run time over 15 minutes.

To our knowledge, no eco-friendly HPLC method was found for separation of seven second-generation antihistamines, with minor modification in mobile phase organic content.

The present study aims to develop a single chromatographic setup for analysis of seven antihistaminic drugs in pharmaceutical formulations. We intended to propose highly efficient and accurate HPLC methods that abide by the green analytical chemistry (GAC) principles. Additionally, a stability-indicating method was developed for separation of azelastine and three related impurities.

2. Experimental

2.1. Reagents, Sample, and Standard

The certified pharmaceutical secondary standard (assigned purity ≥98%) of azelastine HCl, desloratadine, ebastine, fexofenadine HCl, ketotifen, loratadine, and olopatadine HCl were obtained from Sigma-Aldrich (St. Louis, MA, USA).

The azelastine impurity B (1-benzoyl-2-[(4RS)-1-methylhexahydro-1H-azepin-4-yl]diazane), azelastine impurity D (4-(4-Chlorobenzyl) phthalazin-1(2H)-one), and azelastine impurity E (3-(4-chlorobenzilidene) isobenzofuran-1(3H)-one) were obtained from European Pharmacopoeia (EP) Reference Standard Sigma-Aldrich (St. Louis, MA, USA).

The HPLC grade acetonitrile and analytical grade hydrochloric acid (37%) were obtained from Merck KGaA (Darmstadt, Germany). The purified water was obtained from Millipore Milli-Q® Plus (Millipore, Bradford, USA).

The nasal solution containing 0.9 mg mL−1 of azelastine was obtained from a local pharmacy. The Rino-Lastin® (Aché Pharmaceutical Laboratories, Brazil) formulation is stated to contain following excipients: benzalkonium chloride, citric acid, sorbitol, hypromellose, purified water, disodium edetate dihydrate, and sodium phosphate dibasic dodecahydrate.

2.2. Instrumentation and Chromatographic Conditions

A HPLC system consisted of two solvent delivery pumps, an autoinjector fitted with 20 μL loop, an online degasification system, a column thermostat oven, and an UV/VIS photodiode array detector. The output signal was monitored and integrated using CLASS VP® software (Shimadzu Corporation, Japan).

Analytical conditions were optimized using a Luna-RP C18 column (250x4.6 mm, 5 μm). The mobile phase was constituted of acetonitrile and acidified water (pH 2.1) in the following proportions: 15:85, v/v for DESL, 25:75, v/v for KET and OLO, 32:68, v/v for FEX, 35:65, v/v for AZE and LOR, and 45:55, v/v for EB. The mobile phase flow rate was 1.1 mL/min, and UV detections were made at 220 nm (FEX), 254 nm (EB), 271 nm (LOR), 279 nm (DESL), 287 nm (AZE), and 298 nm (KET and OLO). The volume of injection was fixed at 20 μL.

All analyses were done at room temperature, and the column temperature was controlled at 25°C ± 1. The mobile phase was prepared fresh each day, vacuum-filtered through a 0.45 μm Millipore® (HV) hydrophilic membrane.

2.3. Preparation of Standard Solutions

The stock solutions of azelastine HCl reference standard was prepared by transferring mass equivalent to 25 mg of azelastine into a 25 mL volumetric flask. The volume was completed with mobile phase to obtain 1000 μg mL1 of azelastine standard stock solution. In case of impurities B, D, and E, 5 mg mass was solubilized in 5 mL of the mobile phase, in separate flasks. All subsequent dilutions were made with mobile phase to obtain target concentrations for analysis and method validation.

The stock solutions of DESL, EB, FEX, KET, LOR, and OLO reference standards were prepared likewise, to obtain 1000 μg mL1 of each reference standard, in separate flasks.

2.4. Preparation of Sample Stock Solutions

Five AZE nasal solution preparations (claimed 0.9 mg mL−1 of AZE) were pooled into a single flask. An aliquot equivalent to 10 mg of AZE was transferred to a 50 mL volumetric flask. The volume was completed with mobile phase to obtain 200.0 μg mL−1 of AZE. An ultrasonic water bath was used for 5 min, to aid extraction and solubilization. The resultant sample stock solutions were diluted accordingly and were used to determine the precision and accuracy of the method.

2.5. System Suitability

According to official guidelines, the system suitability tests are an integral part of method development [2022]. Once optimum chromatographic conditions were obtained, a systematic chromatographic analysis of seven antihistaminic drugs was performed. We evaluated capacity factor, retention factors, repeatability, peak asymmetry, and number of theoretical plates. During the process, most appropriate chromatographic conditions were achieved, with acceptable relative standard deviations (RSD), for seven antihistamines, individually.

2.6. Validation of the Prototype HPLC-UV Method for Azelastine HCl

To demonstrate adequateness of the proposed methods for seven antihistaminic drugs, azelastine HCl was selected as prototype drug. The proposed method was fully validated according to international guidelines on validation of a chromatographic method for pharmaceutical drug products and applied in the analysis of AZE in nasal solutions [2124].

The selectivity of the proposed method was evaluated through the analysis of reference standards, sample, and placebo solutions. Besides, peak purity index of obtained chromatograms was accessed by UV diode array detector.

Linearity. The mean peak areas (n=3) were plotted against respective concentrations in a range from 10.0 to 60.0 μg mL−1 of AZE. The linear regression of the curve was accessed by the least-squares method, at five distinct concentration levels. All solutions were prepared fresh, and dilutions were made with mobile phase (pH=2.1).where is the peak area; is concentration; is the slope of the curve, and is intercept of the curve on y-axes.

Precision. The intra- and interday precision (repeatability) were evaluated by repeated analysis of sample solutions (n=10). Fresh sample solutions containing approximately 40.0 μg mL−1 of AZE were analyzed, separately, on three consecutive days. The intra- and interday precisions are expressed as relative standard deviation (RSD) amongst analytical responses [21, 22].

Accuracy. The accuracy was determined by recovery of the standard from the sample matrix. The AZE sample solutions were spiked with known amounts of standard solutions, at three concentration levels, in separate flasks. The resultant fortified sample solutions were analyzed by the proposed method. The recovery amount (R) was calculated by comparing the theoretical and found amounts of AZE standards [23].

Detection Limit (DL) and Quantitation Limits (QL). DL and QL were estimated based on the mean standard deviation and the slope of the AZE calibration curve. The estimated values for DL (3 times signal/noise ratio) and QL (10 times signal/noise ratio) were cross-checked by actual analysis of standard solutions.

Robustness. The robustness of the proposed method was evaluated by the deliberate modification of chromatographic conditions such as organic content, mobile phase pH, and flow rate. The impact of these changes on the AZE peak area, retention time (), and tailing factor () was analyzed. The chromatographic conditions were modified, individually and separately. The AZE sample solutions were injected into the system (n=3) under univariate conditions and the results were statistically analyzed.

3. Results and Discussion

The discovery of third-generation antihistamine drugs has brought additional analytical challenges, besides the therapeutic benefits. However, there is limited information on the related impurities and their probable human cardiotoxic effects and pharmacokinetic and prolonged pharmacodynamics profiles [25, 26].

The third-generation H1 antihistamines have an atypical characteristic in their molecular structures. Depending on the pH of the medium, they often are electrically neutral due to opposite charge at different nonadjacent atoms and/or amphoteric due to the presence of acid and basic groups in their molecular structures [17, 27]. Due to the zwitterionic and/or amphoteric characteristics of these drugs, they often present additional analytical challenges, such as excessive retention and peak tailing due to secondary interactions with the stationary phase [2831]. The low solubility of these drugs, in reverse phase mode, often impacts adversely on chromatographic separations.

Besides these important aspects, it is necessary to constantly seek fast and alternative methods that increase productivity, reduced solvent, and sample consumption. Most of the solvents used in liquid chromatography are hazardous to analysts and environment due to inherent volatility, flammability, and toxicity of these solvents [18]. The high sensitivity is desirable, and GAC methods are welcome with trace quantities of organic waste [18, 27, 3133].

3.1. Method Development and Optimization

We tested several reverse-phase columns, such as conventional C18 and core-shell technology C18 stationary phase, besides, -C8, -CN, and –NH2 columns. The preliminary chromatographic parameters obtained with the RP C-18 LUNA (250x4.6mm, 5 μm) column were satisfactory. This column was selected for optimization of chromatographic conditions for separation of seven antihistaminic drugs and system suitability tests.

The separations obtained with acetonitrile, as organic phase modifier, were better than those obtained with methanol. Neat improvement in peak symmetry and intensity with acetonitrile can be attributed to the solubility of antihistaminic drugs in acetonitrile [17].

According to the pH dependent solubility curves described in the literature [17] and in-lab experimental work, the solubility of all seven antihistaminic drugs shows remarkable improvement in acidic media (ionic form). For example, AZE is moderately soluble in water at neutral pH (0.5 mg mL−1); however at pH 1.7 the value may reach 381.9 mg mL−1 [17]. Likewise, a significant improvement in the solubility of DESL, EB, FEX, KET, LOR, and OLO was found at pH ≤ 2.0.

To improve column efficiency and avoid peak tailing during chromatographic runs, in reverse phase mode, it is essential to assure a single ionic form of molecules [29, 31]. The dissociation constants of AZE, DESL, EB, FEX, KET, LOR, and OLO are described in Table 1 [17]. A mobile phase with pH 2.1 was selected to assure near-total ionized forms (more polar microspecies) of cited antihistamines, throughout chromatographic runs. Further, the selected pH may facilitate drug solubilization and mass transfer of selected antihistaminic drugs during chromatographic interactions, in reverse phase mode.

Finally, most of the commercial reverse phase columns are stable at pH between 2.0 and 8.0. Thus, a mobile phase composed of acetonitrile and acidified water (pH 2.1) was selected for further method development and optimization of chromatographic conditions.

3.2. System Suitability Test

The seven antihistaminic drugs selected for study are structurally related and belong to the same therapeutic class; however chromatographic interaction with the column is dissimilar. To develop a unique eco-friendly method, factors such as mobile phase organic content, flow rate, and volume of injection were optimized through univariate experimental design approach for seven antihistaminic drugs.

The optimum conditions for AZE, DESL, EB, FEX, KET, LOR, and OLO were fixed based on method sensitivity (peak area), efficiency (theoretical plates), asymmetry factor (As), and rapid elution (retention factor). Briefly, standard reference solutions were prepared (200 μg mL−1) separately and injected into the system. The organic content of the mobile phase was optimized for rapid elution of antihistamines, separately. In the following step, mobile phase flow rate was adjusted (0.8, 0.9, 1.0, 1.1, and 1.2 mL.min−1). The impact of injection volume and mobile phase flow rate on peak areas and asymmetry is shown in Figures 14. Finally, injection volume was optimized (2, 5, 10, and 20 μL). Fully optimized conditions for AZE, DESL, EB, FEX, KET, LOR, and OLO are presented in Table 2.


DrugMobile Phase
ACN:H2O (pH 2.1)
Flow variation
(0.8-1.2 mL min−1)
Injection volume variation
(2 – 20 μL)
Detection
(nm)
Retention factor
(k’)
Peak Asymmetry
(As)
Theoretical plates
(N/m)

Azelastine HCl35:65 (v/v)1.0202871.521.3414390

Desloratadine15:85 (v/v)1.0202751.261.3215052

Ebastine45:55 (v/v)1.0202541.541.6012698

Fexofenidine HCl32:68 (v/v)1.0202171.101.0414720

Ketotifen25:75 (v/v)1.0202960.821.7912512

Loratadine35:65 (v/v)1.0202711.261.3815680

Olopatadine HCl25:75 (v/v)1.0202961.631.3315412

Optimum value for each drug.
N/m: theoretical plates per meter.

The detection wavelength for AZE, DESL, EB, FEX, KET, LOR, and OLO was selected based on the DAD detector, and the column temperature was maintained at 25°C ± 1.

Figures 511 show representative chromatograms of AZE, DESL, EB, FEX, KET, LOR, and OLO. All separations were obtained using a RP C-18 LUNA (150x4.6mm, 5 μm) column. The mobile phase consisted of acetonitrile and acidified water (pH 2.1) in the following proportions: 15:85, v/v for desloratadine, 25:75, v/v for ketotifen and olopatadine, 32:68, v/v for fexofenadine, 35:65, v/v for azelastine and loratadine, and 45:55, v/v for ebastine. The flow rate was 1.0 ml min−1 and UV detections were made at 220 nm (FEX), 254 nm (EB), 271 nm (LOR), 279 nm (DESL), 287 nm (AZE), and 298 nm (KET and OLO).

3.3. Evaluation of Greenness of the Proposed Methods

As stated earlier, separation of seven antihistaminic drugs was far better with acetonitrile as compared to other organic solvents tested. Thus, we estimated the greenness of the proposed methods based on the GAC principles, described in the literature [18, 19, 32, 34]. A semiquantitative eco-scale was used to evaluate the greenness of proposed methods. Out of twelve GAC principles, five critical parameters dedicated to analytical methods were used and penalty points were assigned for each adverse factor. Table 3 shows total eco-scale scores for proposed methods. Since the eco-scale value is larger than 75, the proposed methods can be considered excellent in terms of greenness [18, 19].


FactorPenalty points

Reagents (HCl)8

Acetonitrile4

Instruments (HPLC)1

Occupational hazards3

Waste8

Total penalty points24

Analytical Eco-Scale total score76

3.4. Method Validation for AZE

Selectivity. The excipients of the AZE nasal solution do not interfere in the main peak. Besides, the chromatographic peak purity index of AZE standard and sample solutions were evaluated based on photodiode array detector. The three-point peak purity index of AZE reference standard and sample solutions were identical (≥ 0.99). The results support the specificity of the proposed method.

Linearity. The proposed method was found to be linear in the concentration range from 10.0 to 60.0 μg.mL1 of AZE. There was a linear correlation amongst AZE concentrations and corresponding peak area (R2 = 0.9960); besides the standard deviation of the residuals was 1.4%. Relevant data on linearly of proposed methods is presented in Table 4.


STATISTICAL PARAMETERS

Concentration range (μg mL−1)10.0 – 60.0

Slope of the curve ()34754

Intercept ()91113

Correlation coefficient (R2)0.9959

Standard deviation of the residuals1.4%

DL (μg mL−1)1.10

QL (μg mL−1)3.32 (RSD < 2%)

DL= detection limit; QL= quantitation limit.

Precision. The method precision was accessed by repeatability (intra-day) and as intermediate precision (interday) tests. The AZE sample solutions (40.0 μg mL−1) were prepared fresh and injected into the system (n=10), on three consecutive days. Table 5 shows found amounts with respective variations (RSD). The RSD amongst results was below 2%. The obtained results suggest a good precision for the determination of AZE [23, 24, 35].


Accessed AZE (μg mL−1)

Inter-day38.11 ± 0.58
 RSD (%)1.52
Intra-day
 Day 138.43 ± 0.54
 RSD (%)1.40
 Day 238.26 ± 0.42
 RSD (%)1.10
 Day 337.71 ± 0.54
 RSD (%)1.44

The recovery of the proposed method was accessed at three concentration levels (low, intermediate, and high). Table 6 shows acceptable recovery of azelastine reference standard from the sample matrix. The results show a good accuracy of the proposed method [23].


SampleAdded concentration
(μg mL−1)
Found concentration
(μg mL−1)
Recovery
SampleStandardStandard(%)

AZE15.0010.09.31 ± 0.1293.10
15.0025.024.49 ± 0.2297.97
15.0040.039.31 ± 0.1898.27

Detection and Quantitation Limits. The estimated values of DL and QL were 1.10 and 3.32 μg mL−1, respectively (Table 4). The QL was cross-checked by actual analysis of standard solution equivalent to 3.3 μg mL−1 (n=3). The RSD was below 1.62%, far below recommended values [21, 24, 35].

The robustness of proposed method for analysis of AZE was evaluated by deliberate change in mobile phase flow rate (0.9, 1.0, and 1.1 mL mL−1), injection volume (18, 20, and 22uL), and mobile phase organic content (ACN 33, 35 e 37%). There was an insignificant impact on peak retention, peak area, and peak asymmetry (Figures 14).

4. Chromatographic Method for Separation of Azelastine Impurities

The chromatographic separation of azelastine and impurities B, D, and E was obtained using the Luna C18 column (150 × 4.6mm x 5 μm). The mobile phase was composed of solution A (ACN with 0.1% formic acid) and solution B (purified water with 0.1% formic acid). The gradient system was 25 to 75% of solution A in 15 min. The mobile phase flow rate was kept constant at 1.0 mL min−1, and UV detections were made at 228 nm. The volume of injection was fixed at 20 μL. All analyses were done at room temperature, and the column temperature was controlled at 25°C ± 1.

The optimum conditions for AZE and related impurities B, D, and E were fixed based on official system suitability parameters, peak asymmetry <2.0, and peak resolution > 2.0 [7, 12, 21]. The AZE nasal solution sample (250 μg mL−1) was spiked with AZE impurities B, D, and E (10 μg mL−1, each) and injected into the system.

The system suitability parameters are described in Table 7, and representative chromatographic separation of AZE and related impurities B, D, and E is presented in Figure 12.


Retention factor
(k’)
Peak resolution (Rs)Peak asymmetry
(As)

Impurity B-0.351.36

Azelastine0.439.0 (B-Aze)1.36

Impurity D3.0417.14 (Aze-D)1.14

Impurity E5.0413.14 (D-E)1.05

k’: relative retention (retention factor).
Rs: resolution between adjacent peaks.

The selectivity of the method was accessed by peak purity (>0.99) and relative retention (k’) of impurity B (-0.35), AZE B (0.43), impurity D (3.04), and impurity E (5.04) standard solutions, separately.

The system suitability parameters such as peak resolutions between adjacent peaks and peak symmetry were according to the specifications for separation of azelastine and its related impurities described in official compendiums [7, 12, 16].

5. Conclusions

The proposed ecologically correct methods may contribute in high-throughput analysis of seven antihistaminic drugs, with total organic residue < 2.5 mL/analysis. The proposed method for separation of azelastine and related organic impurities B, D, and E is highly efficient and economical. The total analysis time was 15 minutes and the mobile phase is free of ion pairing agents and buffer salts. All separations were obtained on a conventional reverse phase HPLC systems and can be considered excellent in terms of greenness. The proposed methods may subsidize in the evaluation of quality, efficacy, and safety of antihistaminic drugs.

Data Availability

The data used to support the findings of this study are available from the corresponding author upon request.

Conflicts of Interest

The authors declare that there are no conflicts of interest regarding the publication of this paper.

Acknowledgments

This work was supported by the São Paulo Research Foundation (FAPESP), Brazil, under Grant no. 2015/11210-1.

References

  1. R. Pawankar, G. Canonica, S. Holgate, and R. Lockey, “WAO, white book on allergy,” Pediatriya, vol. 5, pp. 55–58, 2012. View at: Publisher Site | Google Scholar
  2. J. Bousquet, R. Dahl, and N. Khaltaev, “GARD (Global Alliance against chronic Respiratory Diseases),” Revue des Maladies Respiratoires, vol. 23, no. 4, pp. 73–75, 2006. View at: Publisher Site | Google Scholar
  3. BRASIL. Ministério da Saúde., DATASUS, (2011). http://www2.datasus.gov.br.
  4. F. E. R. Simons, “Drug therapy: advances in H1-antihistamines,” The New England Journal of Medicine, vol. 351, no. 21, pp. 2203–2217, 2004. View at: Publisher Site | Google Scholar
  5. F. E. R. Simons and F. Winnipeg, “H 1 -Antihistamines?: More relevant than ever in the treatment of allergic disorders,” Clinical Immunology, vol. 112, pp. 42–52, 2003. View at: Publisher Site | Google Scholar
  6. J. W. Slater, A. D. Zechnich, and D. G. Haxby, “Second-generation antihistamines,” Drugs, vol. 57, no. 1, pp. 31–47, 1999. View at: Publisher Site | Google Scholar
  7. The British Pharmacopoeia Commission (MHRA), “Azelastine HCl monograph,” in Br. Pharmacopoiea, The Stationery Office, London, UK, 2018. View at: Google Scholar
  8. T. B. P. C. (MHRA), “Desloratadine monograph,” in Pharmacopoeia, Her Majestys Stationary Office, London, UK, 2018. View at: Google Scholar
  9. The British Pharmacopoeia Commission (MHRA), “Ebastine monograph,” in Pharmacopoeia, Her Majesty’s Stationary Office, London, UK, 2016. View at: Google Scholar
  10. T. B. P. C. (MHRA), “Fexofenadine monograph,” in Pharmacopoeia, Her Majestys Stationary Office, London, UK, 2018. View at: Google Scholar
  11. The British Pharmacopoeia Commission (MHRA), “Loratadine monograph,” in Pharmacopoeia, Her Majesty’s Stationary Office, London, UK, 2018. View at: Google Scholar
  12. The United States Pharmacopeial Convention, “Azelastine HCl monograph,” in United States Pharmacop, pp. 419-420, The United States Pharmacopeial Convention, Rockville, MD, USA, 1st edition, 2018. View at: Google Scholar
  13. The United States Pharmacopeial Convention, “Desloratadine monograph,” in United States Pharmacop. (USP41, NF36), pp. 1177-1178, The United States Pharmacopeial Convention, Rockville, MD, USA, 41st edition, 2018. View at: Google Scholar
  14. The United States Pharmacopeial Convention, “Loradadine monograph,” in United States Pharmacop. (USP41, NF36), pp. 2462–2464, The United States Pharmacopeial Convention, Rockville, MD, USA, 41st edition, 2018. View at: Google Scholar
  15. The United States Pharmacopeial Convention, “Olopatadine HCl monograph,” in United States Pharmacop. (USP41, NF36), pp. 3013-3014, The United States Pharmacopeial Convention, Rockville, MD, USA, 41st edition, 2018. View at: Google Scholar
  16. European Pharmacopoeia Convention, “Azelastine hydrochloride,” in Eur. Pharmacopoeia, pp. 1780-1781, Council of Europe (EDQM), 9th edition, 2018. View at: Google Scholar
  17. Chemicalize, Anti-histaminic drugs, ChemAxon Ltd. (2018). https://chemicalize.com.
  18. H. Shaaban, “New insights into liquid chromatography for more eco-friendly analysis of pharmaceuticals,” Analytical and Bioanalytical Chemistry, vol. 408, no. 25, pp. 6929–6944, 2016. View at: Publisher Site | Google Scholar
  19. A. Gałuszka, Z. M. Migaszewski, P. Konieczka, and J. Namieśnik, “Analytical Eco-Scale for assessing the greenness of analytical procedures,” TrAC Trends in Analytical Chemistry, vol. 37, pp. 61–72, 2012. View at: Publisher Site | Google Scholar
  20. The United States Pharmacopeial Convention, “General chapter, <621> chromatography,” in United States Pharmacop. (USP40, NF35), The United, Rockville, MD, USA, 2017. View at: Google Scholar
  21. The United States Pharmacopeial Convention, “Validation of compendial Methods,” in United States Pharmacop. (USP40, NF35), United States Pharmacopeial Comission, Rockville, MD, USA, 2017. View at: Google Scholar
  22. ICH, ICH Topic Q2 (R1) Validation of Analytical Procedures: Text and Methodology, Int. Conf. Harmon. 1994 (2005) 17.
  23. W. Horwitz and G. W. Latimer, “AOAC Official methods of analysis,” in Off. Methods Anal, Association of Official Analytical Chemists, Gaithersburg, MD, 16th edition, 2005. View at: Google Scholar
  24. BRASIL, Ministério da Saúde. RDC No 166, de 24 de julho de 2017. Dispõe sobre a validação de métodos analíticos e dá outras providências., ANVISA, Brasília, 2017.
  25. L. L. Brunton, B. A. Chabner, and B. C. Knollmann, The Pharmacological Basis of Therapeutics, 2011.
  26. CAS, SciFinder, Chem. Abstr. Serv. (2015). https://scifinder.cas.org.
  27. R. G. Frare and A. K. Singh, “A critical review of physicochemical properties and analytical methods applied to quantitative determination of ebastine,” Critical Reviews in Analytical Chemistry, vol. 48, no. 2, pp. 102–109, 2018. View at: Publisher Site | Google Scholar
  28. R. Nageswara Rao and V. Nagaraju, “An overview of the recent trends in development of HPLC methods for determination of impurities in drugs,” Journal of Pharmaceutical and Biomedical Analysis, vol. 33, no. 3, pp. 335–377, 2003. View at: Publisher Site | Google Scholar
  29. L. R. Snyder, J. J. Kirkland, and J. W. Dolan, Introduction to Modern Liquid Chromatography, John Wiley & Sons, Inc., Hoboken, NJ, USA, 2009. View at: Publisher Site
  30. N. M. Nogueira, J. M. P. Alves, M. I. R. M. Santoro, and A. K. Singh, “Rapid derivative spectrophotometric and chromatographic methods for assay of loratadine in tablets and syrups,” Latin American Journal of Pharmacy, vol. 29, pp. 325–332, 2010. View at: Google Scholar
  31. V. R. Meyer, Practical High-Performance Liquid Chromatography, John Wiley & Sons, Ltd, Chichester, UK, 2010. View at: Publisher Site
  32. P. T. Anastas, “Green chemistry and the role of analytical methodology development,” Critical Reviews in Analytical Chemistry, vol. 29, no. 3, pp. 167–175, 1999. View at: Publisher Site | Google Scholar
  33. H. Shaaban and T. Górecki, “Current trends in green liquid chromatography for the analysis of pharmaceutically active compounds in the environmental water compartments,” Talanta, vol. 132, pp. 739–752, 2015. View at: Publisher Site | Google Scholar
  34. K. Van Aken, L. Strekowski, and L. Patiny, “EcoScale, a semi-quantitative tool to select an organic preparation based on economical and ecological parameters,” Beilstein Journal of Organic Chemistry, vol. 2, no. 1, pp. 1–7, 2006. View at: Publisher Site | Google Scholar
  35. I. International Conference on Harmonization, Validation of analytical procedures: Text and Methodology. ICH-Q2B, in: Int. Conf. Harmon. Tech. Requir. Regist. Pharm. Hum. Use, 1996: p. 13.

Copyright © 2019 Lucas Maciel da Costa et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


More related articles

1249 Views | 347 Downloads | 0 Citations
 PDF  Download Citation  Citation
 Download other formatsMore
 Order printed copiesOrder

Related articles

We are committed to sharing findings related to COVID-19 as quickly and safely as possible. Any author submitting a COVID-19 paper should notify us at help@hindawi.com to ensure their research is fast-tracked and made available on a preprint server as soon as possible. We will be providing unlimited waivers of publication charges for accepted articles related to COVID-19. Sign up here as a reviewer to help fast-track new submissions.