Research Article

Green Analytical Methods for the Separation of Seven Antihistamines: Application in Separation of Azelastine and Related Impurities in Nasal Solution

Table 4

Linear regression data in the analysis of AZE.

STATISTICAL PARAMETERS

Concentration range (μg mL−1)10.0 – 60.0

Slope of the curve ()34754

Intercept ()91113

Correlation coefficient (R2)0.9959

Standard deviation of the residuals1.4%

DL (μg mL−1)1.10

QL (μg mL−1)3.32 (RSD < 2%)

DL= detection limit; QL= quantitation limit.