Research Article

Low-Vacuum Quadrupole Mass Filter Using a Drift Gas

Table 1

Simulation parameters for low-vacuum mass spectrometry.

ParametersValue

Length of the mass filter (L)80 mm
Radius of the mass filter (r0)5 mm
Ion mass (mi)40 amu
Background gas molecular mass (mg)10 amu
Background pressure (p)——
Drift flow rate at inlet (V0)0 L/min
Initial kinetic energy of ions (Ei)5 eV
DC component of RF voltage (U)85 V
Amplitude of RF voltage (V)1080 V
Frequency of RF voltage (ω)4 MHz