Methyl Esterification Combined with Gas Chromatography-Mass Spectrometry (GC-MS) for Determining the Contents of Lubricant to Evaluate the Compatibility of Chlorinated Butyl Rubber Stoppers with Liposome Injections

Table 1

Recoveries of methyl hexadecanoate and methyl stearate in chlorinated butyl rubber stoppers.

AnalyteNo.Spiked (mg)Sample (mg)Detected (mg)Founded (mg)Recovery (%)Average recovery (%)RSD (%)

Methyl hexadecanoate1-10.79602.04812.79070.742693.2997.332.55
1-20.79362.04822.80830.760195.78
1-30.79752.04832.80390.755694.74
2-10.99502.04803.01220.964296.90
2-20.99202.04803.01590.967997.57
2-30.99692.04813.03130.983298.63
3-11.19392.04783.24621.1985100.38
3-21.19042.04843.21311.164797.84
3-31.19632.04823.25471.2065100.85
Methyl stearate1-11.42852.60633.97521.368995.8396.551.29
1-21.43702.60644.00121.394997.07
1-31.42992.60653.99711.390697.25
2-11.78562.60614.34961.743597.64
2-21.79632.60614.30941.703394.82
2-31.78742.60624.33771.731596.88
3-12.14272.60584.68872.082997.21
3-22.15552.60664.64102.034494.38
3-32.14482.60634.70592.099697.89