Biotelemetric Wireless Intracranial Pressure Monitoring: An In Vitro Study

Table 2

Summary of the measurement results.

Sensor Resonance frequency [MHz]Overall shift in resonance frequency [kHz]Overall phase dip change [degree]Overall reflection phase change [degree]Max. resolution from resonance frequency [mmHg]Max. resolution from impedance phase [mmHg]Max. resolution from reflection phase [mmHg]

A132806.840.852.52.5
B31.27207.510.92.52.52.5