Research Article
Reflector-Backed Antenna for UWB Medical Applications with On-Body Investigations
Table 3
Simulated and measured maximum realized gain of the cavity-backed prototype.
| F (GHz) | 3.75 | 4 | 4.25 | R. gain (dBi) | Sim | Meas | Sim | Meas | Sim | Meas |
| Phi = 0° | 7.26 | 6.61 | 7.62 | 6.39 | 7.09 | 2.6 | Phi = 90° | 5.58 | 5.47 | 5.99 | 4.9 | 5.9 | 4.25 | Theta = 90° | 3.73 | 1.23 | 3.65 | 2.08 | 2.78 | 1.09 |
|
|