Research Article
Imaging Performance of Quantitative Transmission Ultrasound
Table 1
Speed-of-sound accuracy against the speed phantom measured with twin pistons.
| Area | Mean (M/S) | SD (M/S) | Uniformity (%) | Actual (M/S) | Error (M/S) | Error (% actual) | Error (SD) |
| High | 1571.9 | 18.1 | 1.15% | 1566 | 5.9 | 0.37% | 0.33 | Low | 1418.8 | 5.69 | 0.40% | 1406 | 12.8 | 0.91% | 2.25 | Water | 1508.6 | 7.09 | 0.46% | 1510 | 1.4 | 0.09% | 0.20 | Background | 1447.9 | 6.24 | 0.43% | 1445 | 2.9 | 0.02% | 0.46 |
|
|