Research Article

Influence of Breast Implant Surface Finishing on Physicochemical and Mechanical Properties before and after Extreme Degradation Studies

Table 3

Mechanical properties for samples as received and after acidic and basic degradation.

RoughnessSmooth membranesRough membranes
DegradationAs receivedAcidicBasicAs receivedAcidicBasic

Mechanical strength (MPa) at 450% straina6.7 ± 0.57.6 ± 1.34.9 ± 1.65.3 ± 0.67.3 ± 1.26.0 ± 1.5
Mechanical strength (MPa) at rupture9.2 ± 0.98.1 ± 0.95.2 ± 1.97.9 ± 1.58.0 ± 0.56.7 ± 2.7
Deformation (%) at rupture556 ± 61465 ± 48416 ± 66539 ± 119473 ± 51396 ± 147

aAverage of samples that reached 450% deformation.