Research Article

Effect of Recycling Protocol on Mechanical Strength of Used Mini-Implants

Table 1

Comparison of fracture torque among diameters and recycling groups (ANOVA).

Groups1.3 mm1.4 mm1.5 mm1.6 mm1.7 mm
MeanSDMeanSDMeanSDMeanSDMeanSD

CG9.05a0.5312.17b0.3216.45c0.4121.61d0.5127.26e0.88 <0.001
18.94a0.4212.21b0.4816.28c0.7321.41d0.9626.98e1.41
28.01a0.6012.52b0.9715.75c1.1421.78d0.4726.96e1.01
38.04a0.4111.58b0.3915.79c0.8121.17d0.4726.72e0.72

Different letters (e.g., CG row—a, b, c, d, and e) represent significant difference of fracture torque among diameters, defined by Tukey tests.
Same letters (e.g., 1.3 mm column—a, a, a, and a) represent similarity of fracture torque among groups, defined by Tukey tests.