Research Article

Ion-Selective Electrode for Anionic Surfactants Using Hexadecyl Trimethyl Ammonium Bromide-Sodium Dodecylsulfate as an Active Ionophore

Table 2

The life time study of the anionic surfactant selective electrode.

PeriodSlopeLinear rangeDetection limit
(weeks) (mV/decade) (mol L−1) (mol L−1)

1−58.05.0 × 10−6 to 2.5 × 10−32.9 × 10−6
2−58.24.6 × 10−6 to 2.5 × 10−32.9 × 10−6
3−58.44.7 × 10−6 to 2.5 × 10−33.0 × 10−6
4−58.15.2 × 10−6 to 2.5 × 10−33.0 × 10−6
5−58.35.5 × 10−6 to 2.7 × 10−33.4 × 10−6
6−57.95.9 × 10−6 to 1.9 × 10−32.2 × 10−6
7−58.46.4 × 10−6 to 2.2 × 10−33.9 × 10−6
8−58.65.4 × 10−6 to 1.8 × 10−32.6 × 10−6
9−57.87.1 × 10−6 to 1.5 × 10−34.9 × 10−6
10−58.06.7 × 10−6 to 2.8 × 10−33.5 × 10−6
11−57.68.1 × 10−6 to 2.4 × 10−34.3 × 10−6
12−57.47.8 × 10−6 to 3.1 × 10−34.1 × 10−6
13−57.28.2 × 10−6 to 3.4 × 10−35.0 × 10−6
14−54.69.1 × 10−6 to 3.5 × 10−36.1 × 10−6
15−52.11.1 × 10−5 to 2.9 × 10−3 7.6 × 10−6
16−41.62.4 × 10−5 to 3.1 × 10−31.6 × 10−5