Research Article

Utility of Activated Glassy Carbon and Pencil Graphite Electrodes for Voltammetric Determination of Nalbuphine Hydrochloride in Pharmaceutical and Biological Fluids

Table 3

Statistical parameters of pharmaceutical dosage form assay of the investigated drug by the proposed SWV method at PGE and GCE and reported method.

ElectrodeDrug[NP⋅HCl] takenProposed methodOfficial method-test-test
µg mL−1Mean ± SD ()Mean ± SD ()

PGE Nalufin (20 mg mL−1)10100.2 ± 0.8099.41 ± 2.001.451.56
2099.98 ± 1.11100.2 ± 1.321.701.81
30100.5 ± 0.6599.74 ± 1.001.851.96

GCE Nalufin (20 mg mL−1)20100.7 ± 1.1099.41 ± 2.002.571.25
40100.90 ± 0.88100.2 ± 1.322.781.45
6099.98 ± 0.94100.74 ± 1.002.851.23

Tabulated -value at 95% confidence limit = 2.77, , degree of freedom = 4.
Tabulated -value at 95% confidence limit = 6.39, .