Research Article

Development of an In-Field Method for the Detection of Barium in Various Water Samples Using Differential Pulse Anodic Stripping Voltammetry

Table 1

Total Ba concentrations determined in 10 drinking water samples, taken from municipal tap water, bores, a rainwater tank, and bottled water in Western Australia. [a] 1. Bottled water (Mt Franklin brand), 2. Rainwater (tank), 3. Drinking fountain water (ECU campus), 4. Tap water (Wanneroo residence), 5. Filtered tap water (Wanneroo residence), 6. Bore (Wangara), 7. Bore (Bullsbrook) 1, 8. Bore (Bullsbrook) 2, 9. Bore (Craigie), and 10. Bore, (Muchea). The F value calculated for these samples is 0.99034.

Sample [a]ASV Mean%RSDICP-MS% Differ. ASV vs. ICP
(μg/L n=3)(μg/L)

17.591.37.13.34
28.4715.29.55.73
34306.14300.03
44117.294190.94
54201.14200.03
64646.14600.49
775.249.572.02.19
873.92.171.02.03
960.910.760.00.74
1048.716.448.00.76