Validation of a Simple and Robust Liebermann–Burchard Colorimetric Method for the Assay of Cholesterol in Selected Milk Products in Ghana

Table 3

Results showing precision of results from the developed method.

Precision parametersMean absorbance ± SDRSD

Intra-assay precisionOutcome from 10 replicate determinations95.97 ± 1.0161.06%
Triplicate determinations from three different concentrations80%–0.032 mg/ml96.88 ± 0.72210.75%
100%–0.040 mg/ml95.97 ± 1.0161.06%
120%–0.048 mg/ml96.83 ± 0.77730.80%

Inter-assay precisionSame analystDay 194.25 ± 0.96971.03%
Day 296.05 ± 1.5161.58%
Day 393.01 ± 1.3141.41%
Different analystsAnalyst 194.53 ± 0.87700.93%
Analyst 295.91 ± 1.6951.77%
Analyst 393.42 ± 1.3061.40%

Acceptance criteria<2%