Optimisation of Simultaneous Tl-201/Tc-99m Dual Isotope Reconstruction with Monte-Carlo-Based Scatter Correction

Table 6

Contrast evaluation with NCAT phantoms. The table shows myocardium versus defect and myocardium versus left ventricle (LV) contrasts for optimised and unoptimised Tl-201/Tc-99m reconstruction and for pure Tl-201 reconstruction. The results were obtained using all the Monte Carlo simulated NCAT phantoms, and the contrasts are average values of the corresponding female and male phantoms. Only Tl-201 reconstruction results are shown.

DefectReconstructionMyocardium versus defect contrastMyocardium versus LV contrastReconstruction time (min)

ANTOptimised0.620.91<3
Unoptimised0.610.91 <13
Pure Tl-2010.710.96<1

INFOptimised0.800.94<3
Unoptimised0.810.94 <13
Pure Tl-2010.900.97<1

SEPTOptimised0.700.96<3
Unoptimised0.670.95 <13
Pure Tl-2010.640.99<1