Research Article
Feasibility and Initial Performance of Simultaneous SPECT-CT Imaging Using a Commercial Multi-Modality Preclinical Imaging System
Table 3
Peak-to-valley ratios for each profile drawn through the rod phantom.
| 4.8 mm and 4.0 mm rods | 37 min SPECT | % Diff. of mean | 72 min SPECT | % Diff. of mean | SPECT-CT | % Diff. of mean |
| 25.93 | 0.6 | 25.75 | 1.28 | 26.57 | 1.88 |
| 3.2 mm rods | 37 min SPECT | % Diff. of mean | 72 min SPECT | % Diff. of mean | SPECT-CT | % Diff. of mean |
| 15.05 | 7.65 | 12.58 | 9.96 | 14.3 | 2.31 |
| 2.4 mm rods | 37 min SPECT | % Diff. of mean | 72 min SPECT | % Diff. of mean | SPECT-CT | % Diff. of mean |
| 6.23 | 19.51 | 8.65 | 11.85 | 8.33 | 7.65 |
|
|