99mTc-Sestamibi/123I Subtraction SPECT/CT in Parathyroid Scintigraphy: Is Additional Pinhole Imaging Useful?

Table 3

False-negative and false-positive findings with –sestamibi/123I subtraction SPECT/CT alone and with an additional pinhole image. Case number with a letter (a and b) refers to the same patient.

Case numberLocationClinical backgroundSPECT/CTSPECT/CT with pinholeComment

1LLpHPTFNTPResidual activity was faintly visible in SPECT/CT, but was classified as negative. With additional pinhole image correct classification was done.
2ULpHPTFNTP
3aLLsHPTFNTP
4aLLpHPTFNTP

3bLRsHPTFNFNNo clear residual activity
5aURpHPTFNFN
6URpHPTFNFN

7LLpHPTFPTNVery uneven iodine uptake
4bURpHPTFPTNSubtraction artefact (edge effect)
8LLpHPTFPTNSubtraction artefact (edge effect)
9LRpHPTFPTNAn iodine-negative nodule

5bLLpHPTFPFPA large thyroid cyst at left side
10aLRpHPTFPFPSubtraction artefact
10bLLpHPTFPFPSubtraction artefact

UL = upper left; LL = lower left; LR = lower right; UR = upper right.