Optical Path Length Calibration: A Standard Approach for Use in Absorption Cell-Based IR-Spectrometric Gas Analysis

Table 1

Comparison of path lengths derived from the presented OPC ( ), from mechanical measurements ( ), and from gas standard-based retrieval ( ), respectively. Given are normalized errors referring to .

CellCell type /m /m /m Speciesa / m

AMP*21.87 ± 0.04021.26 ± 0.310.9CO22.0
BSP, HP*0.819 ± 0.00290.822 ± 0.00100.50.805 ± 0.0410.2CH41.65
CSP, BW*0.773 ± 0.00200.774 ± 0.00100.20.773 ± 0.0380.0CH41.65
CSP, BW*0.773 ± 0.00200.774 ± 0.00100.20.774 ± 0.0120.04CO22.71
DSP, HT*0.380 ± 0.00120.380 ± 0.00200.00.380 ± 0.0190.0CH41.65

Gas standards probed at wavelengths specified in column.
*Cell characteristics: MP multipass White type, SP single pass cell, HP high pressure cell, BW Brewster windows-equipped, and HT high temperature cell.